首页> 外文期刊>Talanta: The International Journal of Pure and Applied Analytical Chemistry >Online monitoring of trace chlorinated benzenes in flue gas of municipal solid waste incinerator by windowless VUV lamp single photon ionization TOFMS coupled with automatic enrichment system
【24h】

Online monitoring of trace chlorinated benzenes in flue gas of municipal solid waste incinerator by windowless VUV lamp single photon ionization TOFMS coupled with automatic enrichment system

机译:无窗VUV灯单光子电离TOFMS结合自动富集系统在线监测城市垃圾焚烧炉烟气中微量氯苯。

获取原文
获取原文并翻译 | 示例
           

摘要

Chlorinated benzenes are typical precursors and indicators for polychlorinated dibenzo-p-dioxins and dibenzofurans (PCDD/Fs) emissions' rom waste incinerators. Online and real-time monitoring of chlorobenzenes is a challenge due to their low concentration and complex nature of the flue gas. In this work, a continuous online monitoring system was built for detection of trace chlorinated benzenes based on a time-of-flight mass spectrometer (TOFMS). A single photon ionization (SPI) source based on a radiofrequency-excited windowless vacuum ultraviolet (VUV) lamp was developed for the first time to eliminate the signal attenuation resulting from the contamination of magnesium fluoride windows and to avoid the fragment ions. An automatic enrichment system including three parallel Tenax TA adsorption tubes was designed and coupled to the TOFMS to achieve the required ultrahigh sensitivity. The limits of quantitation at 7.65, 5.37 and 6.77 pptv were obtained for monochlorobenzene (MCBz), dichlorobenzene (DCBz) and trichlorobenzene (TrCBz), respectively, within a 29-min analytical period. Moreover, this apparatus was applied to continuously online monitor the actual flue gas from a waste incinerator for three months. During this period, the concentrations of MCBz, DCBz and TrCBz detected in the flue gas were in the range of 100-1200, 50-800 and 50-300 pptv, respectively. The relative standard deviation (RSD) of the sensitivity for the windowless VUV lamp ion source was 9.71% evaluated by the internal standard benzene over the 3-months flue gas monitoring. These results demonstrated the capability of this method in long-term analysis of the trace chlorinated benzenes in the flue gas.
机译:氯化苯是多氯联苯对二恶英和二苯并呋喃(PCDD / Fs)排放垃圾焚烧炉的典型前体和指示剂。由于氯苯浓度低且烟气性质复杂,因此在线和实时监测氯苯是一个挑战。在这项工作中,建立了基于飞行时间质谱仪(TOFMS)的连续在线监测系统,用于检测痕量氯苯。首次开发了基于射频激发的无窗真空紫外(VUV)灯的单光子电离(SPI)源,以消除由于氟化镁窗口污染而导致的信号衰减并避免碎片离子。设计了一个自动浓缩系统,该系统包括三个平行的Tenax TA吸附管,并与TOFMS耦合以实现所需的超高灵敏度。在29分钟的分析时间内,对一氯苯(MCBz),二氯苯(DCBz)和三氯苯(TrCBz)分别获得了7.65、5.37和6.77 pptv的定量限。此外,该设备还用于连续在线监测废物焚烧炉中的实际烟气三个月。在此期间,烟气中检测到的MCBz,DCBz和TrCBz的浓度分别在100-1200、50-800和50-300 pptv的范围内。在3个月的烟气监测中,通过内标苯评估了无窗VUV灯离子源灵敏度的相对标准偏差(RSD)为9.71%。这些结果证明了该方法对烟道气中痕量氯苯的长期分析的能力。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号