This paper presents the first quantitative study of realtime acquisition of images of spots on thin-layer chromatographic plates during development. Procedures are described for imaging using a CCD camera and for image processing, incorporating corrections for fixed pattern effects and compensation for the moving solvent front, to measure the absorbance of the analyte. Imaging of Sudan II was carried out in transmission mode, and peak areas were found to be time-independent. Quantification of the relationship between peak area and sample loading was established over the range 1-50 ng. After averaging 55 images obtained during a single chromatographic run, which attenuates noise contributions from local nonuniformities in the sorbent layer, precision and detection limits were found to be comparable with values obtained in previous work using offline measurements.
机译:本文提出了在开发过程中实时获取薄层色谱板上斑点图像的第一个定量研究。描述了使用CCD相机成像和图像处理的程序,包括对固定图案效果的校正和对移动溶剂前沿的补偿,以测量分析物的吸光度。 Sudan II的成像以透射模式进行,发现峰面积与时间无关。在1-50 ng范围内建立了峰面积与样品上样量之间关系的定量分析。在对单次色谱运行中获得的55张图像进行平均后,这减弱了吸附剂层中局部不均匀性造成的噪声影响,发现精确度和检测限可与之前使用离线测量获得的值相当。