首页> 外文期刊>Analytical chemistry >Results with an apparatus for pressurized planar ectrochromatography
【24h】

Results with an apparatus for pressurized planar ectrochromatography

机译:用加压平面外色谱仪得到的结果

获取原文
获取原文并翻译 | 示例
       

摘要

Pressurized planar electrochromatography (PPEC) is a fast and efficient planar chromatographic technique. The mobile phase is driven by electroosmotic flow, while the system is pressurized in a manner that allows heat to flow between the sorbent layer and the pressurizing medium. The reproducibility of solute retention was not satisfactory in the initial report describing PPEC. In the current report, this reproducibility is improved by better control of several experimental variables. The pressure at which PPEC is performed is now free of drift, and the temperature at which the layer is preconditioned is maintained to within 1 T. The best reproducibility of retention is obtained when the plate is soaked in the mobile phase for a defined time before each run. In the original prototype, the temperature of the sorbent layer was not controlled. In the present apparatus, water, at a constant temperature between 3 and 60 degrees C, is circulated through channels in the two die blocks that pressurize the layer. The highest efficiency is obtained at an intermediate temperature. This behavior is ascribed to high resistance to mass transfer at the lower temperatures and increased diffusion at higher temperatures. Efficiency, as measured by the number of theoretical plates, increases with increasing migration distance. The height equivalent of a theoretical plate diminishes with increasing migration distance, and values as low as 0.0106 mm are obtained under appropriate conditions. This extrapolates to 94 000 plates/m. Manual spotting was used in this report. Evidence is presented that substantially better efficiency would be obtained if the initial spot size were smaller. The efficiency of PPEC in its current form is illustrated by a chromatogram showing the separation of nine solutes in 2 min. PPEC was also performed with TLC plates in a back-to-back configuration, and this doubles the number of samples that can be simultaneously separated.
机译:加压平面电色谱(PPEC)是一种快速而有效的平面色谱技术。流动相由电渗流驱动,而系统以允许热量在吸附剂层和加压介质之间流动的方式加压。在描述PPEC的最初报告中,溶质保留的重现性并不令人满意。在当前报告中,通过更好地控制几个实验变量可以改善这种可重复性。现在执行PPEC的压力没有漂移,并且将层进行预处理的温度保持在1 T以内。当将板在流动相中浸泡规定的时间之前,可获得最佳的保留重现性。每次运行。在原始的原型中,吸附剂层的温度不受控制。在本设备中,使水在3至60摄氏度之间的恒定温度下循环通过两个模块中的通道,从而对层加压。在中间温度下可获得最高效率。这种现象归因于在较低温度下对传质的高抵抗力以及在较高温度下的扩散增加。用理论塔板数衡量的效率随迁移距离的增加而增加。理论平板的高度当量随着迁移距离的增加而减小,并且在适当的条件下可获得低至0.0106 mm的值。外推至94 000板/米。此报告中使用了手动发现。有证据表明,如果初始光斑尺寸较小,则可获得实质上更好的效率。色谱图显示了PPEC当前形式的效率,该色谱图显示了2分钟内分离出的9种溶质。 PPEC还使用TLC板以背对背的配置进行,这使可同时分离的样品数量增加了一倍。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号