Pulsation modes and the effect of applied voltage on current and flow rate in nanoelectrospray

机译:纳米电喷雾的脉动模式和施加电压对电流和流速的影响

获取原文
获取原文并翻译 | 示例
       

摘要

New modes of nanoelectrospray (nES) have been identified that exhibit stable high-frequency current oscillations in the kilohertz range. In-line flow rate measurements during nES have for the first time allowed the accurate determination of the spray current and volumetric flow rate dependence on applied voltage. High-speed video imaging has revealed that each current pulse corresponds to the periodic formation of a short-lived stable cone-jet structure at the capillary tip. These findings are of particular importance and suggest greater process monitoring and control in nES-MS systems is necessary if efficiency and ion yield are to be optimized.
机译:已经确定了纳米电喷雾(nES)的新模式,它们在千赫兹范围内表现出稳定的高频电流振荡。 nES期间的在线流速测量首次允许精确确定喷雾电流和体积流速对所施加电压的依赖性。高速视频成像显示,每个电流脉冲都对应于毛细管尖端处短暂稳定的锥形喷射结构的周期性形成。这些发现特别重要,表明如果要优化效率和离子产率,则需要在nES-MS系统中进行更大的过程监控。