The effect of hydrogel and silicone hydrogel contact lenses on the measurement of intraocular pressure with rebound tonometry

机译:水凝胶和有机硅水凝胶隐形眼镜对回弹眼压计测量眼内压的影响

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

Purpose: To assess the accuracy of intraocular pressure (IOP) measurements using rebound tonometry over disposable hydrogel (etafilcon A) and silicone hydrogel (senofilcon A) contact lenses (CLs) of different powers. Methods: The experimental group comprised 36 subjects (19 male, 17 female). IOP measurements were undertaken on the subject's right eyes in random order using a rebound tonometer (ICare). The CLs had powers of +2.00D, -2.00D and -6.00D. Six measurements were taken over each contact lens and also before and after the CLs had been worn. Results: A good correlation was found between IOP measurements with and without CLs (all r>0.80; p<0.05). Bland Altman plots did not show any significant trend in the difference in IOP readings with and without CLs as a function of IOP value. A two-way ANOVA revealed a significant effect of material and power (p<0.01) but no interaction. All the comparisons between the measurements without CLs and with hydrogel CLs were significant (p < 0.01). The comparisons with silicone hydrogel CLs were not significant. Conclusions: Rebound tonometry can be reliably performed over silicone hydrogel CLs. With hydrogel CLs, the measurements were lower than those without CLs. However, despite the fact that these differences were statistically significant, their clinical significance was minimal.
机译:目的:通过回弹眼压计评估不同屈光度的一次性水凝胶(etafilcon A)和有机硅水凝胶(senofilcon A)的眼压(IOP)测量的准确性。方法:实验组包括36名受试者(男19名,女17名)。使用回弹眼压计(ICare)以随机顺序对受试者的右眼进行IOP测量。 CL的功率为+ 2.00D,-2.00D和-6.00D。在佩戴隐形眼镜之前和之后,对每个隐形眼镜进行六次测量。结果:在有和没有CL的情况下,IOP测量之间发现良好的相关性(所有r> 0.80; p <0.05)。布兰特·奥特曼(Bland Altman)图没有显示有和没有CL的IOP读数随IOP值变化的任何显着趋势。双向方差分析显示物质和能量的显着影响(p <0.01),但没有相互作用。在不使用CL和使用水凝胶CL的情况下,所有测量值之间的比较都是显着的(p <0.01)。与有机硅水凝胶CL的比较不显着。结论:回弹眼压计可以在硅水凝胶CL上可靠地进行。在使用水凝胶CL的情况下,其测量值低于没有使用CL的情况。但是,尽管这些差异具有统计学意义,但其临床意义却很小。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号