A Collimated LED Light Module for Absorbance Detection

机译:用于吸光度检测的准直LED光模块

获取原文
获取原文并翻译 | 示例
           

摘要

In the bio absorbance detection systems, sensitivity of detection depends on collimation length of light module. Although the conventional LEDs have the advantages of low cost and compact, their light is diverged with divergence angle larger than 10 degree. Therefore, conventional LEDs are not suitable for bio-chip absorbance detection. By exploiting the concept of fiber collimator, this project proposes a design to realize a collimated light module. The advantages of low cost, miniaturized, and multi-channel capability are then achieved. The divergence angle of this collimator is about 1.2 degree within 1.5 cm optical path length. The power variation of this module is under 1% within an hour. The absorbance experiment to measure sodium pyruvate range from 0 to 82μM using fabricated light module was conducted, where sensitivity of 7.7×10{sup}3 / M and linearity of 0.9995 R{sup}2 is obtained.
机译:在生物吸收检测系统中,检测的灵敏度取决于光模块的准直长度。尽管常规LED具有低成本和紧凑的优点,但是它们的光以大于10度的发散角发散。因此,常规的LED不适用于生物芯片吸收度检测。通过利用光纤准直仪的概念,该项目提出了一种实现准直光模块的设计。这样便获得了低成本,小型化和多通道功能的优势。该准直仪的发散角在1.5厘米的光路长度内约为1.2度。该模块的功率变化在一小时内低于1%。进行了人工光模块测定丙酮酸钠在0至82μM范围内的吸光度实验,灵敏度为7.7×10 {sup} 3 / M,线性为0.9995 R {sup} 2。