For prompt and precise control of an extracorporeal circulation system, we developed an apparatus which continuously measures the occlusion degree of a roller pump-driven circuit electrically. The principle of our measurement is that the degree of the occlusion of a tube fulfilled with electrically conductive fluid correlates with the electrical resistance between two electrodes set both at inflow and outflow side of the tube occluding site. Our preliminary study using this apparatus elucidated that the correlation between the measured electrical resistance and the degree of the tube occlusion was almost linear. It was also revealed that the electrical resistance at the proper occlusion degree determined by Japan Industrial Standards (JIS) was approximately ten times that at the open, not occluded, state. With our apparatus, more simple and precise application of proper tube occlusion degree was feasible than with the previous method checking the fluid reflux in the occluded tube. Moreover, the continuous measurement of the tube pressure during the roller pump works might detect the deviation of the rotation axis of the roller pump so that early detection of the pump trouble would be possible. In the future, our equipment will enable not only the automated detection of the proper tube occlusion degree but also the automated regulatory system for the extra-corporeal circulation.
机译:为了迅速而精确地控制体外循环系统,我们开发了一种设备,该设备可以连续地电气测量滚筒泵驱动电路的阻塞程度。我们的测量原理是,充满导电流体的管子的堵塞程度与设置在管子堵塞部位的流入侧和流出侧的两个电极之间的电阻相关。我们使用该设备进行的初步研究表明,测得的电阻与管阻塞程度之间的相关性几乎是线性的。还揭示了由日本工业标准(JIS)确定的在适当的闭塞度下的电阻大约是在开放而不闭塞的状态下的电阻的十倍。使用我们的设备,比使用先前的方法检查闭塞管中的流体回流的方法更简单,更准确地应用适当的闭塞度是可行的。而且,在辊筒泵工作期间连续测量管压力可以检测辊筒泵的旋转轴线的偏差,从而可以及早发现泵的故障。将来,我们的设备不仅可以自动检测出适当的管阻塞程度,还可以实现体外循环的自动调节系统。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号