Flow through a mechanical distraction enterogenesis device: a pilot test.

机译:流经机械式分散肠肠发生装置:先导试验。

获取原文
获取原文并翻译 | 示例

摘要

BACKGROUND: We tested the coupling portion of a prototype intraluminal distraction enterogenesis device to allow flow-through of simulated enteric contents (SEC) in both pig and human jejunum. MATERIALS AND METHODS: SEC was made using 80% corn syrup. Ten-cm pig and human intestinal segments had a spoke-shaped 2.2 cm coupling adaptor sutured in place, intraluminally. The adaptor had a flow-through area of 33.6 mm(2). SEC was pumped into the proximal part of the intestinal segment at 0.083 mL/s. The times to first passage of SEC through the coupler (first drop), 10 mL, and 20 mL of SEC eluted from the distal end were recorded. RESULTS: Mean time to first drop elution was 155 +/- 38 s with pig, and 149 +/- 22 s with human bowel (P = 0.8). This corresponded to a hydrostatic pressure of 37.5 mmHg before the initial drop passed through. Mean flow rates were 0.094 mL/s in pig bowel and 0.084 mL/s in human bowel (P = 0.09). To account for occlusion from luminal debris, a 75% occlusion of coupler holes was studied in the smaller pig bowel to investigate if reductions in flow-through area could be tolerated. Mean time to first drop increased slightly to 171 +/- 15 s, but the elution rate stayed the same (P = 0.5). CONCLUSIONS: After a physiologic level of initial pressure buildup allowing the first drop of SEC to pass the coupling adaptor, our prototype intestinal coupling adaptor did not obstruct flow-through of SEC, even after a 75% decrease in flow-through area. This type of attachment represents a viable approach to placing a device in-continuity without obstructing flow of enteric contents.
机译:背景:我们测试了原型腔内分流肠发生装置的耦合部分,以允许猪和人空肠中的模拟肠内容物(SEC)流通。材料与方法:SEC用80%玉米糖浆制成。十厘米的猪和人的肠段在腔内缝合有辐条状的2.2厘米耦合适配器。适配器的流通面积为33.6 mm(2)。将SEC以0.083 mL / s的速度泵入肠段的近端。记录SEC首次通过耦合器(第一滴)的时间,从远端洗脱的10 mL和20 mL SEC。结果:猪的第一次滴落平均时间为155 +/- 38 s,人肠为149 +/- 22 s(P = 0.8)。这对应于初始滴通过之前的37.5 mmHg的静水压力。猪肠的平均流速为0.094 mL / s,人肠的平均流速为0.084 mL / s(P = 0.09)。为了解决管腔碎片的阻塞问题,在较小的猪肠中研究了75%的耦合孔阻塞问题,以研究是否可以忍受流通面积的减少。第一次下降的平均时间略微增加到171 +/- 15 s,但洗脱速率保持不变(P = 0.5)。结论:在生理水平的初始压力建立,使SEC的第一滴通过耦合适配器后,即使在流通面积减少了75%之后,我们的原型肠耦合适配器也没有阻碍SEC流通。这种类型的附件表示一种可行的方法,可在不阻碍肠内容物流动的情况下将设备连续放置。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号