The performance of a recently developed real-time PCR system, the ExiStation HBV diagnostic system, for quantitation of hepatitis B virus (HBV) in human blood was evaluated. The detection limit, reproducibility, cross-reactivity, and interference were evaluated as measures of analytical performance. For the comparison study, 100 HBV-positive blood samples and 100 HBV-negative samples from Korean Blood Bank Serum were used, and the results of the ExiStation HBV system showed good correlation with those obtained using the Cobas TaqMan (r(2) = 0.9931) and Abbott real-time PCR systems (r(2) = 0.9894). The lower limit of detection was measured as 9.55 IU/mL using WHO standards and the dynamic range was linear from 6.68 to 6.68 x 10(9) IU/mL using cloned plasmids. The within-run coefficient of variation (CV) was 9.4%, 2.1%, and 1.1%, and the total CV was 11.8%, 3.6%, and 1.7% at a concentration of 1.92 log(10) IU/mL, 3.88 log(10) IU/mL, and 6.84 log(10) IU/mL, respectively. No cross-reactivity or interference was detected. The ExiStation HBV diagnostic system showed satisfactory analytical sensitivity, excellent reproducibility, no cross-reactivity, no interference, and high agreement with the Cobas TaqMan and Abbott real-time PCR systems, and is therefore a useful tool for the detection and monitoring of HBV infection. (C) 2013 Elsevier B.V. All rights reserved.
机译:评估了最近开发的实时PCR系统ExiStation HBV诊断系统对人血中乙型肝炎病毒(HBV)定量的性能。评价检测限,重现性,交叉反应性和干扰作为分析性能的指标。为了进行比较研究,使用了韩国血库血清中的100例HBV阳性血液样本和100例HBV阴性样本,并且ExiStation HBV系统的结果与使用Cobas TaqMan获得的结果具有良好的相关性(r(2)= 0.9931 )和雅培实时PCR系统(r(2)= 0.9894)。使用WHO标准,检测下限为9.55 IU / mL,使用克隆质粒动态范围从6.68至6.68 x 10(9)IU / mL呈线性。在1.92 log(10)IU / mL,3.88 log的浓度下,批内变异系数(CV)为9.4%,2.1%和1.1%,总CV为11.8%,3.6%和1.7% (10)IU / mL和6.84 log(10)IU / mL。未检测到交叉反应或干扰。 ExiStation HBV诊断系统表现出令人满意的分析灵敏度,出色的重现性,无交叉反应,无干扰,并且与Cobas TaqMan和Abbott实时PCR系统高度吻合,因此是检测和监测HBV感染的有用工具。 (C)2013 Elsevier B.V.保留所有权利。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号