首页> 外文期刊>Journal of vascular and interventional radiology: JVIR >Thrombogenicity of various endovascular stent types: an in vitro evaluation.
【24h】

Thrombogenicity of various endovascular stent types: an in vitro evaluation.

机译:各种血管内支架类型的血栓形成性:体外评估。

获取原文
获取原文并翻译 | 示例
获取外文期刊封面目录资料

摘要

PURPOSE: The aim of this study was to evaluate the thrombogenicity of different peripheral stent types in a standardized in vitro model with fresh human whole blood. MATERIALS AND METHODS: Different stents (N = 77; n = 7 of each of 11 types) were implanted in polyvinyl chloride tubing loops and filled with donor blood samples. After 120 minutes of blood circulation, the thrombin-antithrombin III complex (TAT) levels, beta-thromboglobulin (beta-TG) levels, and platelet counts were assessed. RESULTS: After 2 hours, significant differences were seen. TAT values (+/- SD) with the investigated stents were 31 micro g/mL +/- 20 (control, no stent), 328 micro g/mL +/- 206 (Saxx stent, peripheral medium CrNi31 L), 651 micro g/mL +/- 760 (Palmaz Corinthian Stent, 316 L stainless steel, electropolished), 1,609 micro g/mL +/- 1,264 (Palmaz Corinthian Stent, 316 L stainless steel, not electropolished), 810 micro g/mL +/- 578 (Palmaz Schatz long medium stent), 569 micro g/mL +/- 347 (Smart Nitinol stent), 1,037 micro g/mL +/- 577 (Megalink peripheral stent), 543 micro g/mL +/- 487 (peripheral stent, electropolished), 1,674 micro g/mL +/- 2,057 (peripheral stent, not electropolished), 3,128 micro g/mL +/- 1,812 (SelfX Nitinol stent, polished), 5,897 micro g/mL +/- 2,380 (SelfX Nitinol stent, unpolished), and 1,458 micro g/mL +/- 887 (bridge stent). The platelet count (x1,000/ micro L +/- SD) was 218 +/- 35 (control, no stent), 188 +/- 22 (Saxx stent), 187 +/- 20 (Palmaz Corinthian stent, electropolished), 135 +/- 37 (Palmaz Corinthian stent, not electropolished), 170 +/- 24 (Palmaz Schatz stent), 180 +/- 36 (Smart Nitinol stent), 159 +/- 26 (Megalink peripheral stent), 173 +/- 17 (peripheral stent, electropolished), 133 +/- 51 (peripheral stent, not electropolished), 123 +/- 37 (SelfX Nitinol stent, polished), 52 +/- 27 (SelfX Nitinol stent, unpolished), and 130 +/- 31 (bridge stent). CONCLUSION: This standardized study showed a wide range of platelet activation after stent implantation. Electropolishing clearly reduced the thrombogenicity of the stents.
机译:目的:本研究的目的是在新鲜人全血的标准化体外模型中评估不同外周支架类型的血栓形成性。材料与方法:将不同的支架(N = 77; 11种类型中的每种,n = 7)植入聚氯乙烯管路中,并充满供血样本。血液循环120分钟后,评估凝血酶-抗凝血酶III复合物(TAT)水平,β-凝血球蛋白(β-TG)水平和血小板计数。结果:2小时后,观察到显着差异。使用研究的支架的TAT值(+/- SD)为31 micro g / mL +/- 20(对照,无支架),328 micro g / mL +/- 206(Saxx支架,周围介质CrNi31 L),651 micro g / mL +/- 760(Palmaz Corinthian支架,316 L不锈钢,电抛光),1,609 micro g / mL +/- 1,264(Palmaz Corinthian支架,316 L不锈钢,未电抛光),810 micro g / mL + / -578(Palmaz Schatz长中等支架),569 micro g / mL +/- 347(Smart Nitinol支架),1,037 micro g / mL +/- 577(Megalink外围支架),543 micro g / mL +/- 487(外围支架,电抛光),1,674 micro g / mL +/- 2,057(外围支架,未电抛光),3,128 micro g / mL +/- 1,812(SelfX Nitinol支架,抛光),5,897 micro g / mL +/- 2,380( SelfX Nitinol支架(未抛光)和1,458 micro g / mL +/- 887(桥支架)。血小板计数(x1,000 / micro L +/- SD)为218 +/- 35(对照,无支架),188 +/- 22(Saxx支架),187 +/- 20(Palmaz Corinthian支架,电抛光) ,135 +/- 37(Palmaz Corinthian支架,未电抛光),170 +/- 24(Palmaz Schatz支架),180 +/- 36(Smart Nitinol支架),159 +/- 26(Megalink外围支架),173 + /-17(外围支架,电抛光),133 +/- 51(外围支架,未电抛光),123 +/- 37(SelfX Nitinol支架,抛光),52 +/- 27(SelfX Nitinol支架,未抛光)和130 +/- 31(桥接支架)。结论:这项标准化的研究表明支架植入后血小板活化范围广泛。电抛光明显降低了支架的血栓形成性。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号