Significant caval penetration by the celect inferior vena cava filter: attributable to filter design?

机译:折衷的下腔静脉滤器对腔的穿透力强:可归因于滤器设计?

获取原文
获取原文并翻译 | 示例
           

摘要

This report describes transmural penetration of the inferior vena cava (IVC) by a newly introduced IVC filter within 9 days of its placement. A computed tomographic study demonstrated filter penetration with one of the primary struts lodging in the uncinate process of the pancreas. Because of the close resemblance of this new filter to another filter that has not been associated with penetration injuries, the key differences between the two designs were examined and the determination was made that the new filter, unlike the older one, has unprotected primary struts. Only filters with an unprotected primary strut design have been associated with penetration injuries such as the one described in this case.
机译:本报告介绍了新引入的IVC过滤器在放置后9天内对下腔静脉(IVC)的透壁穿透性。一项计算机断层扫描研究表明,滤过器的穿透作用是在胰腺的未融合过程中出现了一个主要的支撑物。由于此新过滤器与未发生渗透损伤的另一过滤器非常相似,因此对这两种设计之间的主要区别进行了检查,并确定新过滤器与旧过滤器不同,其主支柱没有保护。只有具有不受保护的主支撑杆设计的过滤器才与穿透伤相关,例如本例中所述。