首页> 外文期刊>Journal of vascular and interventional radiology: JVIR >Bench-top evaluation of air flow through a valved peelable introducer sheath.
【24h】

Bench-top evaluation of air flow through a valved peelable introducer sheath.

机译:台式评估通过阀式可剥皮导引器护套的气流。

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

PURPOSE: Air embolism is a rare but potentially fatal complication that may occur during the insertion of a central venous catheter. A valved peelable introducer sheath was developed to reduce the likelihood of an air embolus. This study was performed to determine the rate of air flow through this valved introducer sheath under different conditions that may be encountered in a clinical setting. MATERIALS AND METHODS: A 16-F-diameter valved peelable introducer sheath was used for all experiments. A bench-top testing system was constructed that consisted of a vacuum source, a mass flow meter, and a digital vacuum gauge. A coupling device with a pneumatic O-ring was used to seal and connect the distal end of the introducer sheath to the testing system. A vacuum of -5 mm Hg was applied to the sealed distal end of the sheath to simulate physiologic conditions. The rate of air flow through the valved sheath was measured under three different conditions; (i) valve open, (ii) valve closed, and (iii) valve openbut with the sheath manually pinched. Thirty air flow measurements were performed for each of the three test conditions. RESULTS: When the valve was in the open position, the mean rate of air flow through the introducer sheath was 417.2 mL/sec (range, 415.5-419.7 mL/sec). When the valve was in the closed position, the mean rate of air flow was 0.004 mL/sec (range, 0.000-0.067 mL/sec). When the valve was open but the sheath was manually pinched, the mean rate of air flow through the sheath was 31.7 mL/sec (range, 23.0-38.8 mL/sec). During the 90 testing procedures, the mean vacuum was -5.10 mm Hg (range, -5.00 to -5.45 mm Hg). CONCLUSIONS: This bench-top study revealed that a massive amount of air flowed through the open 16-F introducer sheath when it was subjected to a vacuum of -5 mm Hg, a situation that may occur under normal physiologic conditions. The rate of air flow could be decreased by aggressively pinching the sheath but the rate of air flow was still substantial. This study demonstrated that the addition of a sophisticated valve mechanism can essentially eliminate air flow through a peelable introducer sheath.
机译:目的:空气栓塞是一种罕见的但可能致命的并发症,可能会在中央静脉导管插入过程中发生。开发了带阀的可剥皮导引器护套,以减少出现空气栓塞的可能性。进行这项研究是为了确定在临床条件下可能遇到的不同条件下,通过有瓣导尿管鞘的气流速率。材料与方法:所有实验均使用直径为16-F的带阀可剥式导引鞘管。构建了一个台式测试系统,该系统由真空源,质量流量计和数字真空计组成。使用带有气动O形圈的连接装置将导引器护套的远端密封并连接到测试系统。将-5 mm Hg的真空施加到护套的密封远端,以模拟生理状况。在三种不同条件下测量了通过带阀护套的空气流量。 (i)阀门打开,(ii)阀门关闭,和(iii)阀门打开,但护套被手动捏紧。对于三个测试条件中的每一个,进行了三十次气流测量。结果:当阀门处于打开位置时,通过导引鞘的平均空气流速为417.2 mL / sec(范围415.5-419.7 mL / sec)。当阀处于关闭位置时,平均气流速率为0.004 mL / sec(范围为0.000-0.067 mL / sec)。当阀门打开但手动捏紧护套时,通过护套的平均空气流量为31.7 mL / sec(范围为23.0-38.8 mL / sec)。在90个测试过程中,平均真空度为-5.10毫米汞柱(范围-5.00至-5.45毫米汞柱)。结论:这项台式研究表明,当真空度为-5 mm Hg时,大量空气流经开放的16-F导引鞘,在正常的生理条件下可能会发生这种情况。可以通过用力挤压护套来降低气流速率,但是气流速率仍然很高。这项研究表明,添加先进的阀门机构可以从根本上消除通过可剥皮导引器护套的气流。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号