首页> 外文期刊>Journal of vascular and interventional radiology: JVIR >Measurement of portal blood and transjugular intrahepatic portosystemic shunt flow with use of a retrograde thermodilutional catheter.
【24h】

Measurement of portal blood and transjugular intrahepatic portosystemic shunt flow with use of a retrograde thermodilutional catheter.

机译:使用逆行热稀释导管测量门静脉血和经颈静脉肝内门体分流。

获取原文
获取原文并翻译 | 示例
获取外文期刊封面目录资料

摘要

PURPOSE: Direct blood flow measurement in the portal vein (PV) and in transjugular intrahepatic portosystemic shunts (TIPS) holds potential as a means of optimizing shunt creation and predicting outcomes. A retrograde thermodilutional flow measurement catheter was recently designed. The purpose of this study was to validate and optimize the technique of blood flow measurement in the PV with use of this catheter in a swine model of TIPS. MATERIALS AND METHODS: Six domestic swine underwent TIPS creation. At day 14, shunts were catheterized for venography and portal blood flow measurement with use of thermodilutional catheters. Simultaneous flow measurements were made with use of the thermodilutional and perivascular methods at the same anatomic location within the PV. To test reproducibility of the flow measurement, repeated measurements were performed in the PV and shunt. RESULTS: Fifty-five measurements of PV blood flow with the thermodilutional catheter and perivascular probe were made in five animals. Perivascular probe placement was unsuccessful in one animal. There was high correlation between the thermodilutional and perivascular measurements (r(2) = 0.96; Q(periv) = 0.80 . Q(therm) + 147 mL/min; n = 55; range, 205-1541 mL/min). The difference between two consecutive thermodilutional measurements and their mean was less than 50 mL/min in 72% of measurements and less than 100 mL/min in 94% of measurements (N = 85 pairs). CONCLUSIONS: Measurement of PV blood flow with use of a retrograde thermodilutional catheter system yields reliable and reproducible results. This catheter system holds promising utility as a device for measuring flow within TIPS and the PV.
机译:目的:直接测量门静脉(PV)和经颈静脉肝内门体分流术(TIPS)的血流,有望成为优化分流术产生和预测预后的手段。最近设计了逆行热稀释流量测量导管。这项研究的目的是验证和优化在TIPS猪模型中使用该导管的PV中血流测量技术。材料与方法:六只家猪进行了TIPS创作。在第14天,使用热稀释导管对分流器进行静脉造影和门脉血流量测量。使用热稀释法和血管周方法在PV内的相同解剖位置同时进行流量测量。为了测试流量测量的可重复性,在PV和分流器中进行了重复测量。结果:用热稀释导管和血管周探针对五只动物进行了55次PV血流测量。一只动物的血管周围探针放置失败。热稀释和血管周围测量之间具有高度相关性(r(2)= 0.96; Q(periv)= 0.80。Q(therm)+ 147 mL / min; n = 55;范围205-1541 mL / min)。两次连续的热稀释法测量值之间的差值在72%的测量值中均小于50 mL / min,在94%的测量值中均小于100 mL / min(N = 85对)。结论:使用逆行热稀释导管系统测量PV血流可产生可靠且可重复的结果。该导管系统作为一种用于测量TIPS和PV内流量的装置具有广阔的应用前景。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号