Highly cross-linked polyethylene acetabular liners retrieved four to five years after revision surgery: a report of two cases.

机译:在翻修手术后四到五年就可以找到高度交联的聚乙烯髋臼衬里:两例报道。

获取原文
获取原文并翻译 | 示例

摘要

There is currently considerable interest in the use of highly cross-linked polyethylene (XLPE) acetabular liners for total hip arthroplasty (THA). In literature, only a single retrieval analysis of one type of XLPE liner implanted for greater than four years exists. The purpose of the present report is to quantify surface deviations in two XLPE liners implanted during revision THA and retrieved between four to five years after implantation. The two XLPE acetabular liners (Reflection, Smith and Nephew Inc., Memphis, TN) were retrieved from patients undergoing their second revision surgery, at 4.90 and 4.07 years. The retrieved liners and a new, non-implanted, unworn liner of the same size were scanned using micro-computed tomography (micro-CT). Articular surface deviation maps were created by comparing the retrievals to the unworn liner, based on the liner geometry obtained from micro-CT. The linear penetration rates were found to be 0.018 and 0.008 mm/year. Localized scratches and pits with deviations greater than 0.205 mm were also found on the articular surfaces of both liners. The XLPE liners retrieved from the two cases demonstrated low linear penetration rates. Regions with greater focal deviations were also apparent, likely due to third-body wear. The results are consistent with previously published clinical follow-ups of other XLPE liners.
机译:当前,高度交联的聚乙烯(XLPE)髋臼内衬用于全髋关节置换术(THA)的使用引起了极大的兴趣。在文献中,仅存在对植入四年以上的一种XLPE衬里的单次检索分析。本报告的目的是量化THA修订版中植入的两根XLPE衬垫的表面偏差,并在植入后四到五年之间恢复。从接受第二次翻修手术的患者(分别为4.90岁和4.07岁)取回了两个XLPE髋臼内衬(Reflection,Smith和Nephew Inc.,田纳西州孟菲斯)。使用微型计算机断层扫描(micro-CT)扫描取回的衬管和尺寸相同的新的未植入未磨损衬管。关节表面偏差图是根据从微型CT获得的衬管几何形状,通过将取回物与未磨损的衬管进行比较而创建的。发现线性渗透率是0.018和0.008mm /年。在两个衬套的关节表面上也发现了偏差大于0.205毫米的局部划痕和凹坑。从这两个案例中获得的XLPE衬管显示出较低的线性渗透率。也可能出现焦距偏差较大的区域,这可能是由于第三身体磨损所致。结果与其他XLPE衬管先前发表的临床随访结果一致。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号