Development and Validation of RP-HPLC Method for Determination of Famotidine and its Application in Quality Control of Different Pharmaceutical Dosage Forms

机译:RP-HPLC法莫替丁测定方法的建立与验证及其在不同剂型质量控制中的应用

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

A precise and fast novel high-performance liquid chromatography method was developed and validated for the quantitative determination of Famotidine (FMT) in commercially available pharmaceutical dosage forms. An Agilent 1200 Series High Performance Liquid Chromatography (HPLC) system having the column C_(18) (5μm particle size, 150x4.6 mm) was used in this study and detection (diode array detector) was made at 280 nm. The mobile phase was acetonitrile, distilled water, triethylamine and phosphoric acid (49.9:49.9:0.1:0.1, v/v), isocratic elution under ambient temperature at flow rate of 1.5 mL min~(-1) with injection volume 5μL, In this method, the retention times for FMT pure, tablets and suspension were 0.787 min, 0.789 min and 0.839 minutes respectively. The new method was validated by different validation parameters. The procedure provided a linear response over the concentration range of 0.1-1.0 mg mL~(-1) (r~2 =0.998) and equation was y=3902.6+18.651. The mean % recovery for inter-day (96.56%) and intra-day (97.36%) assuring a good precision and accuracy was 96-98%. The method was found to be very rapid and the overall assay time was less than 2 minutes and the results obtained were accurate, precise and selective enough to allow the determination of FMT in the presence of certain excipients.
机译:开发了一种精确快速的新型高效液相色谱方法,并经过验证可用于定量测定市售药物剂型中的法莫替丁(FMT)。在本研究中,使用具有C_(18)柱(5μm粒径,150x4.6 mm)的Agilent 1200系列高效液相色谱(HPLC)系统,并在280 nm处进行检测(二极管阵列检测器)。流动相为乙腈,蒸馏水,三乙胺和磷酸(49.9:49.9:0.1:0.1,v / v),在环境温度下以1.5 mL min〜(-1)的流速进行等度洗脱,进样量5μL,In该方法对FMT纯,片剂和悬浮液的保留时间分别为0.787分钟,0.789分钟和0.839分钟。通过不同的验证参数对新方法进行了验证。该程序在0.1-1.0 mg mL〜(-1)的浓度范围内提供线性响应(r〜2 = 0.998),方程为y = 3902.6 + 18.651。日间(96.56%)和日间(97.36%)的平均回收率(确保良好的准确性和准确性)为96-98%。发现该方法非常快速,总测定时间少于2分钟,并且获得的结果足够准确,精确和选择性足以在某些赋形剂存在下测定FMT。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号