首页> 外文期刊>Journal of the Brazilian Chemical Society >Graphite electrodes modified by 8-hydroxyquinolines and its application for the determination of copper in trace levels
【24h】

Graphite electrodes modified by 8-hydroxyquinolines and its application for the determination of copper in trace levels

机译:8-羟基喹啉修饰的石墨电极及其在痕量铜测定中的应用

获取原文
获取原文并翻译 | 示例
获取外文期刊封面目录资料

摘要

Surface modification by 8-hydroxyquinoline-5-sulfonic acid (8-HQS) or 8-hydroxyquinoline (8-HQ) on a graphite electrode through irreversible adsorption is reported in this paper. Cyclic voltammetry was used to characterize the surface behavior. The modified surface exhibited an affinity to chelating Cu(II) in the solution, forming a Cu(II) complex, which was employed for Cu(II) trace analysis. Of the metals Zn, Ni, Pb, Co, and Cd, none presented interference until excess concentration of 10 times. Significant interference could be observed from Co(II), Cd(II) and Fe(II) for an excess concentration of 100 times on the analyte. A differential pulse voltammetry, combined with a preconcentrating-stripping process and a standard addition method was used for the analysis. A detection limit for trace copper determination in water, such as 5.1 x 10(-9) mol L-1, was obtained.
机译:本文报道了通过不可逆吸附作用在石墨电极上用8-羟基喹啉-5-磺酸(8-HQS)或8-羟基喹啉(8-HQ)进行表面改性的方法。循环伏安法用于表征表面行为。修饰的表面表现出对溶液中螯合Cu(II)的亲和力,形成Cu(II)络合物,可用于Cu(II)痕量分析。在Zn,Ni,Pb,Co和Cd等金属中,直到浓度超过10倍时才出现干扰。在分析物上浓度超过100倍时,Co(II),Cd(II)和Fe(II)可能会出现明显的干扰。差分脉冲伏安法结合预浓缩-汽提过程和标准添加方法用于分析。获得了用于水中痕量铜测定的检测极限,例如5.1 x 10(-9)mol L-1。