The effects of receiver placement on probe microphone, performance, and subjective measures with open canal hearing instruments.

机译:接收器放置对开放式管道助听器的探头麦克风,性能和主观测量的影响。

获取原文
获取原文并翻译 | 示例
           

摘要

BACKGROUND: Open canal hearing instruments differ in method of sound delivery to the ear canal, distance between the microphone and the receiver, and physical size of the devices. Moreover, RITA (receiver in the aid) and RITE (receiver in the ear) hearing instruments may also differ in terms of retention and comfort as well as ease of use and care for certain individuals. What remains unclear, however, is if any or all of the abovementioned factors contribute to hearing aid outcome. PURPOSE: To determine the effect of receiver location on performance and/or preference of listeners using open canal hearing instruments. RESEARCH DESIGN: An experimental study in which subjects were exposed to a repeated measures design. STUDY SAMPLE: Twenty-five adult listeners with mild sloping to moderately severe sensorineural hearing loss (mean age 67 yr). DATA COLLECTION AND ANALYSIS: Participants completed two six-week trial periods for each device type. Probe microphone, objective, and subjective measures (quiet, noise) were conducted unaided and aided at the end of each trial period. RESULTS: Occlusion effect results were not significantly different between the RITA and RITE instruments; however, frequency range was extended in the RITE instruments, resulting in significantly greater maximum gain for the RITE instruments than the RITA instruments at 4000 and 6000 Hz. Objective performance in quiet or in noise was unaffected by receiver location. Subjective measures revealed significantly greater satisfaction ratings for the RITE than for the RITA instruments. Similarly, preference in quiet and overall preference were significantly greater for the RITE than for the RITA instruments. CONCLUSIONS: Although no occlusion differences were noted between instruments, the RITE did demonstrate a significant difference in reserve gain before feedback at 4000 and 6000 Hz. Objectively; no positive benefit was noted between unaided and aided conditions on speech recognition tests. These results suggest that such testing may not be sensitive enough to determine aided benefit with open canal instruments. However, the subjective measures (Abbreviated Profile of Hearing Aid Benefit [APHAB] and subjective ratings) did indicate aided benefit for both instruments when compared to unaided. This further suggests the clinical importance of subjective measures as a way to measure aided benefit of open-fit devices.
机译:背景技术:开放式耳道助听器在将声音传递到耳道,麦克风与接收器之间的距离以及设备的物理尺寸方面有所不同。此外,RITA(辅助设备中的接收器)和RITE(耳朵中的接收器)助听器在保持性和舒适性以及对某些人的易于使用和护理方面也可能有所不同。但是,尚不清楚上述因素中的任何一个或全部是否有助于助听器结果。目的:确定使用开放式管道助听器的接收器位置对听众的表现和/或偏好的影响。研究设计:一项实验研究,对象被接受重复测量的设计。研究样本:25名成年听众,轻度倾斜至中度严重的感音神经性听力损失(平均年龄67岁)。数据收集和分析:参与者针对每种设备类型完成了两个为期六周的试用期。在每个试验期结束时,均未进行探针麦克风,客观和主观措施(安静,噪音)的测试和帮助。结果:RITA和RITE仪器的遮挡效果结果无显着差异。但是,RITE仪器的频率范围得到了扩展,导致RITE仪器在4000和6000 Hz时的最大增益明显大于RITA仪器。安静或嘈杂环境中的客观表现不受接收机位置的影响。主观测量表明,RITE的满意度远高于RITA仪器。同样,相对于RITA仪器,RITE对安静和整体的偏爱要大得多。结论:尽管在仪器之间没有发现咬合差异,但是RITE确实显示了在4000和6000 Hz反馈之前储备增益的显着差异。客观地在语音识别测试的独立条件和辅助条件之间没有发现任何积极的好处。这些结果表明,此类测试可能不够灵敏,无法确定开放式运河器械的辅助益处。但是,与无助相比,主观衡量指标(助听器收益的简要说明[APHAB]和主观评分)确实表明了两种工具的助益。这进一步表明,主观测量作为一种测量开放式设备辅助收益的方法的临床重要性。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号