Assessment of on-screen measurements, magnification, and calibration in digital radiography

机译:评估数字射线照相术中的屏幕测量,放大倍率和校准

获取原文
获取原文并翻译 | 示例
           

摘要

Objective—To compare calibration methods for digital radiography in terms of measurement accuracy and interobserver variability.Design—Prospective study.Sample—Digital radiographic images of a 155-mm-long Steinmann pin.Procedures—Measurement of pin length on digital radiographs was determined with a 25.4-mm-diameter calibration ball and commercially available software program via 3 calibration methods (ie, no calibration, autocalibration, and manual calibration). Digital radiographs of the calibration ball and pin were obtained with each placed at various vertical heights from the table (7 heights) and horizontal distances from the center of the beam (4 distances). Measurements of pin length on digital radiographs were made by 4 observers who were blinded to the orientation of the calibration ball and pin.Results—Pin lengths obtained by each calibration method were significantly different from each other and from the true value. Manual calibration was the most accurate. There was no significant interobserver variability in measurements. There was no significant change in measurements when the calibration ball was moved horizontally, but pin length measurements changed significantly when the ball was moved vertically (away from the table) with an approximate magnification error of 1 % per centimeter of distance between the calibration ball and pin.Conclusions and Clinical Relevance—For digital radiography, manual calibration is recommended to achieve the most accurate measurements. Ideally, the calibration ball should be placed at the same vertical height as the object to be measured; however, ifthis cannot be achieved, the magnification error can be expected to be approximately 1 % per centimeter of distance.
机译:目的-比较数字射线照相的校准方法在测量精度和观察者之间的差异性。设计-前瞻性研究。样品-155毫米长的Steinmann针的数字射线照相图像。程序-用数字射线照相确定针长度的测量一个直径25.4毫米的校准球和可通过3种校准方法(即无校准,自动校准和手动校准)购买的软件程序。获得校准球和销的数字X射线照片,每张照片都放置在距桌子不同的垂直高度(7个高度)和距光束中心的水平距离(4个距离)上。由4位观察员进行的数字X射线照片上销钉长度的测量,他们对校准球和销钉的方向视而不见。结果-通过每种校准方法获得的销钉长度彼此之间以及与真实值之间均存在显着差异。手动校准是最准确的。测量中观察者间无显着差异。水平移动校准球时,测量值没有显着变化,但是垂直移动(远离工作台)时,针长测量值发生了显着变化,校准球与测量球之间的距离的厘米误差约为1%。针。结论和临床意义-对于数字X射线照相术,建议进行手动校准以实现最准确的测量。理想情况下,校准球应放置在与被测物体相同的垂直高度上;但是,如果无法做到这一点,则可以预期放大误差约为每厘米距离1%。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号