首页> 外文期刊>Journal of radiological protection: Official journal of the Society for Radiological Protection >Proposals for revised factors in intra-oral radiography shielding calculations
【24h】

Proposals for revised factors in intra-oral radiography shielding calculations

机译:关于口内射线照相屏蔽计算中修正因素的提案

获取原文
获取原文并翻译 | 示例
       

摘要

The current value for weighted average primary plus scatter kerma at 1m for intra-oral radiography shielding assessment is based on scatter radiation measurements made with the trunk of a RANDO phantom and an assessment of primary transmission that is unverified. Measurements of primary transmission and scatter radiation during intra-oral radiography were made at 30° intervals through a full 360° rotation using two anthropomorphic head phantoms and similar equipment at three different sites. The results suggest that a scatter factor of 5Gy(Gycm ~2) ~(-1) and a primary transmission of 0.03% of the entrance surface dose are more appropriate and, therefore, we recommend that the weighted average primary plus scatter kerma used for shielding calculations can be reduced from 1 to 0.5Gy per exposure at 1m. This factor will adequately account for exposures made at 60 and 70kV using a range of intra-oral units.
机译:口内射线照相术屏蔽评估中1m的加权平均初级和散布比释动能的当前值基于对RANDO体模躯干进行的散布辐射测量以及未经验证的初级透射评估。使用两个拟人化的头部模型和三个不同位置的类似设备,以360°的旋转角度在30°间隔内进行口内X射线摄影期间一次透射和散射辐射的测量。结果表明,散度因子为5Gy(Gycm〜2)〜(-1)和初次透射率为入射表面剂量的0.03%更合适,因此,我们建议将加权平均初次加散射比释比用于屏蔽计算可以从1m每次曝光的1Gy降低到0.5Gy。该因素将充分考虑使用一系列口内单位在60kV和70kV下进行的暴露。