AutoCal: A method for automatic energy calibration of β-γ Systems based on a 137Cs spectrum

机译:AutoCal:一种基于137Cs光谱的β-γ系统自动能量校准方法

获取原文
获取原文并翻译 | 示例
           

摘要

A method was developed and implemented that calculates energy versus channel calibrations. This method utilizes a ~(137)Cs spectrum acquired in β-γ coincidence. Both the γ singles and the β-γ coincident spectrum are utilized. The γ singles spectrum is utilized to calculate the γ detector energy versus resolution function. The Compton line in the β-γ coincidence is then used to calculate the β detector energy versus resolution function. Currently a linear regression (y = mx + b) is utilized to fit the data. However, higher order polynomials may easily be implemented if desired.
机译:开发并实现了一种计算能量与通道校准的方法。该方法利用以β-γ重合获得的〜(137)Cs光谱。 γ单峰和β-γ重合谱都被利用。 γ单谱用于计算γ检测器能量与分辨率的函数。然后使用β-γ重合中的康普顿线计算β检测器能量与分辨率的函数。目前,线性回归(y = mx + b)用于拟合数据。但是,如果需要,可以容易地实现高阶多项式。