PURPOSE: To describe a series of myopic patients in whom a LASIK flap was created with the IntraLase femtosecond laser. The LADARVision (Alcon Laboratories Inc) tracking system failed and resulted in postponement of surgery. METHODS: Five myopic Asian patients with darkly pigmented irides underwent bilateral superior hinge flap creation with three different femtosecond laser (IntraLase Corp) pulse frequencies: 15, 30, and 60 kHz. Mean patient age, preoperative best spectacle-corrected visual acuity (BSCVA), and spherical equivalent refraction were 33 years, 20/20, and -6.25 diopters (D), respectively. Despite uncomplicated flap-lift and adequate pharmacologic dilated pupil diameter >7 mm, difficulty was experienced with LADARVision eye tracking in 9 of 10 eyes. Intraoperative manipulations (fluid application, levitation of the laser bed, and tracking through a repositioned flap) were attempted to facilitate eye tracking at the time of femtosecond laser flap creation, but without success. RESULTS: The flap was repositioned, and patients were asked to return 1 to 5 days later, at which time the flap was relifted, eye tracking was successfully achieved, and the surgery was completed. Mean postoperative uncorrected visual acuity was 20/18 and no eyes lost any lines of BSCVA at 1 year. CONCLUSIONS: Irregularities of the IntraLASIK stromal bed surface induce forward light scatter, and dark ocular pigmentation further reduces retinal reflectance, leading to tracking failure with the Alcon LADARVision platform. Delaying the surgery by 1 to 5 days may aid in reducing the light scatter, so that eye tracking can be achieved.
机译:目的:描述一系列近视患者,其中使用IntraLase飞秒激光创建了LASIK皮瓣。 LADARVision(Alcon Laboratories Inc)跟踪系统发生故障,导致手术推迟。方法:五名近视眼的亚洲黑色虹膜患者使用三种不同的飞秒激光(IntraLase Corp)的脉冲频率分别为15、30和60 kHz进行了双侧上皮瓣的制作。平均患者年龄,术前最佳眼镜矫正视力(BSCVA)和球面等效屈光度分别为33岁,20/20和-6.25屈光度(D)。尽管皮瓣抬高不复杂且药理学扩张瞳孔直径大于7 mm,但在10眼中有9眼的LADARVision眼动追踪中遇到了困难。试图进行术中操作(流体施加,悬浮激光床以及通过重新定位的皮瓣进行跟踪)来促进飞秒激光皮瓣创建时的眼睛跟踪,但没有成功。结果:皮瓣被重新定位,并要求患者在1-5天后返回,此时皮瓣被再次提起,成功地完成了眼动追踪,并完成了手术。术后1年的平均未矫正视力为20/18,并且没有眼睛失去任何BSCVA系。结论:IntraLASIK基质床表面的不规则性会引起前向光散射,而深色眼色素沉着会进一步降低视网膜反射率,从而导致Alcon LADARVision平台跟踪失败。将手术延迟1-5天可以帮助减少光散射,从而可以实现眼睛跟踪。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号