Precision of NIDEK OPD-scan measurements.

机译:NIDEK OPD扫描测量的精度。

获取原文
获取原文并翻译 | 示例
获取外文期刊封面目录资料

摘要

PURPOSE: To evaluate the repeatability of wavefront measurements using the NIDEK OPD-Scan. METHODS: A total of 179 eyes from 90 healthy volunteers (57 women and 33 men) with no corneal or lenticular pathology and normal visual acuity were enrolled in this study. Mean patient age was 39 years (range: 17 to 85 years). All patients underwent four consecutive measurements by one examiner with the NIDEK OPD-Scan. Total, corneal, and internal wavefront errors were measured and calculated with the device, using slit retinoscopy. Repeatability of the measurements was evaluated for spherical aberration, coma, and trefoil. RESULTS: The repeatability test revealed a good result for all three higher order aberrations evaluated. The best repeatability values were found for total aberrations, followed by internal and corneal aberrations. CONCLUSIONS: The NIDEK OPD-Scan has good precision in the wavefront measurement of total, corneal, and internal optical aberrations.
机译:目的:评估使用NIDEK OPD-Scan进行波前测量的可重复性。方法:本研究共纳入了来自90名健康志愿者(179名女性,分别为57名女性和33名男性)的179只眼睛,这些患者没有角膜或晶状体病理,并且视力正常。平均患者年龄为39岁(范围:17至85岁)。由一名检查员使用NIDEK OPD-Scan对所有患者进行了四次连续测量。使用狭缝检影法通过该设备测量和计算总,角膜和内部波前误差。针对球差,彗差和三叶形评估了测量的可重复性。结果:重复性测试显示,对所有三个更高阶像差进行了评估,均获得了良好的结果。对于总像差,发现最佳的重复性值,其次是内部和角膜像差。结论:NIDEK OPD-Scan在波前测量总,角膜和内部光学像差方面具有很高的精度。