Test-retest and inter-rater reliability of a method to measure wrist and finger spasticity

机译:重测和评定者间可靠性的一种测量手腕和手指痉挛的方法

获取原文
获取原文并翻译 | 示例
       

摘要

Objective: To describe the reliability of an instrument (Neuro Flexor) designed to be used in the clinic for quantification of the relative contribution of spasticity, elasticity and viscosity to resistance during passive wrist movements. Design: A test-retest and inter-rater reliability study. Subjects: A convenience sample of 34 adults with chronic stroke with spasticity in the hand, and a reference group of 10 healthy persons. Methods: Two raters assessed the participants with the NeuroFlexor. Elastic, viscous and neural components of passive movement resistance were quantified at the wrist. Test-retest and inter-rater intraclass correlation coefficient (ICC 2.1) were calculated for each component for both raters and two sessions. Degree of measurement error was evaluated using the coefficient of variation and the repeatability coefficient. Results: Reliability was high for the neural component (testretest: 0.90-0.96; inter-rater: 0.90-0.94), fair to good for the elastic component (test-retest: 0.79-0.88; inter-rater: 0.76-0.76), and fair to high for the viscous component (testretest: 0.88-0.90; inter-rater: 0.75-0.80). Based on test-retest data, the coefficients of variation for the neural, elastic and viscous components were 25%, 26% and 16%, respectively, and the repeatability coefficients were 1.798, 1.897 and 1.404, respectively. Conclusion: The NeuroFlexor instrument is a reliable measure of spasticity and of muscle elasticity and viscosity in individuals with wrist and finger muscle resistance to passive stretch after stroke.
机译:目的:描述一种设计用于临床的仪器(神经弯曲仪)的可靠性,该仪器用于量化被动腕部运动过程中痉挛性,弹性和粘性对抵抗力的相对贡献。设计:重新测试和评估者间的可靠性研究。受试者:一组方便样本,其中34例患有中风且患有痉挛的慢性中风成年人,以及10名健康人的参考组。方法:两名评估者使用NeuroFlexor评估了参与者。被动运动阻力的弹性,粘性和神经成分在腕部定量。为评估者和两个会话的每个组件计算了测试-重新测试和评估者间类内相关系数(ICC 2.1)。使用变化系数和重复性系数评估测量误差的程度。结果:神经组件的可靠性较高(测试重新测试:0.90-0.96;中间评定者:0.90-0.94),弹性组件的可靠性相当好(重新测试:0.79-0.88;中间评定者:0.76-0.76),粘性成分的测试结果从中到高(testretest:0.88-0.90; inter-rater:0.75-0.80)。根据重测数据,神经,弹性和粘性成分的变异系数分别为25%,26%和16%,重复性系数分别为1.798、1.897和1.404。结论:NeuroFlexor仪器是对手腕和手指肌肉对中风后被动拉伸有抵抗力的人的痉挛性以及肌肉弹性和粘度的可靠度量。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号