首页> 外文期刊>Journal of liquid chromatography and related technologies >Trace Level Determination of Cr(III)/Cr(VI) in Water Samples Using Ion Chromatography with UV Detection
【24h】

Trace Level Determination of Cr(III)/Cr(VI) in Water Samples Using Ion Chromatography with UV Detection

机译:离子色谱-紫外检测法测定水样中痕量Cr(III)/ Cr(VI)

获取原文
获取原文并翻译 | 示例
           

摘要

In the present research, a method of simultaneous determination of Cr(III) and Cr(IV) with mixed bed ion-exchange column was carried out. Cr(III) is oxidized to Cr(VI) in post-column derivatization reaction, then both ions are detected by a UV detector, at wavelength A = 365 nm.The performance characteristics of the method were established by determining the following validation parameters: precision and accuracy, linearity, limit of detection, and limit of quantification. The influence of sample pH and selected inorganic ions on chromium species analyses has been checked. The method was applied for simultaneous determination of Cr(III)/Cr(VI) in environmental samples such as, rainwater and galvanic sediments.
机译:本研究采用混合床离子交换柱同时测定Cr(III)和Cr(IV)的方法。在柱后衍生化反应中将Cr(III)氧化为Cr(VI),然后通过紫外检测器在波长A = 365 nm处检测到两个离子,通过确定以下验证参数来建立该方法的性能特征:精度和准确性,线性,检测限和定量限。已经检查了样品的pH值和选定的无机离子对铬种类分析的影响。该方法适用于同时测定雨水和电沉积物等环境样品中的Cr(III)/ Cr(VI)。