Effect of engaging abutment position in implant-borne, screw-retained three-unit fixed cantilevered prostheses.

机译:种植体固定的,螺钉固定的三单元固定悬臂假体的接合基台位置的影响。

获取原文
获取原文并翻译 | 示例
           

摘要

PURPOSE: To investigate the effects of internally connected engaging component position in screw-retained fixed cantilevered prostheses. MATERIALS AND METHODS: Twenty-one three-unit fixed dental prostheses (FDPs) were cast in high-palladium alloy in three groups. In group A, engaging components were incorporated into the units away from the cantilevered segment; proximal units received nonengaging components. In group B, these positions were reversed. Control specimens were fabricated using all nonengaging components. Specimens were attached to internally connected 3.5 (diameter) x 13 mm (length) implants, torqued to 32 Ncm, and embedded into epoxy resin. Specimens were tested in cyclic fatigue with a 2 Hz sine wave and 0.1 min/max load ratio. Load amplitude started at 1.8 N and increased by 1.8 N every 60 cycles until fracture. Log-rank statistic, ANOVA, Spearman's correlation, and LIFETEST procedures were used to evaluate level of statistical significance within the results. RESULTS: In the control group, the mean number of cycles to fracture was 31,205 +/- 2639. Mean axial force at fracture was 932 +/- 78 N. In group A, these numbers were 38,160 +/- 4292 and 1138 +/- 128 N, and in group B, 31,810 +/- 3408 and 949 +/- 101 N. Statistical significance levels for number of cycles to fracture were: Control versus group A, p = 0.0117, and groups A versus B, p = 0.0156 (statistically significant). Control versus group B, p = 0.357 (not statistically significant). Log-rank statistic for the survival curves is greater than would be expected by chance; there was a statistically significant difference between survival curves (p = 0.012). The location and mode of failure were noteworthy (always in the abutment screw). CONCLUSIONS: The position of the engaging component had significant effects on the results. Within the limitations of this investigation, it can be concluded that using an engaging abutment in a screw-retained fixed cantilevered FDP provides a mechanical advantage, and engaging the implant furthest from the cantilever when designing a screw-retained cantilever FDP increased resistance to fracture of the distal abutment screw.
机译:目的:研究内部连接的接合部件位置在螺钉固定的固定悬臂假体中的影响。材料与方法:在高钯合金中分三组铸造二十一个三单元固定义齿(FDP)。在A组中,将接合部件合并到远离悬臂梁的单元中;近端单元接受非接合组件。在B组中,这些位置相反。使用所有非接合组件制作对照样品。将样品附着到内部连接的3.5(直径)×13 mm(长度)的植入物上,拧紧至32 Ncm,并嵌入环氧树脂中。在2 Hz正弦波和0.1 min / max负载比的循环疲劳下测试样品。负载振幅从1.8 N开始,每60个周期增加1.8 N,直到断裂。对数统计,ANOVA,Spearman相关和LIFETEST程序用于评估结果中的统计显着性水平。结果:在对照组中,平均断裂循环数为31205 +/-2639。在断裂时的平均轴向力为932 +/- 78N。在A组中,这些数字为38160 +/- 4292和1138 + / -128 N,在B组中为31,810 +/- 3408和949 +/- 101N。断裂循环数的统计学显着性水平为:对照组与A组,p = 0.0117,以及A与B组,p = 0.0156(具有统计意义)。对照组与B组相比,p = 0.357(无统计学意义)。生存曲线的对数秩统计大于偶然预期。生存曲线之间存在统计学差异(p = 0.012)。失效的位置和方式是值得注意的(总是在基台螺丝中)。结论:接合部件的位置对结果有重要影响。在本研究的范围内,可以得出结论,在螺纹固定的固定悬臂FDP中使用接合基台具有机械优势,在设计螺纹固定的悬臂FDP时,将离悬臂最远的植入物接合起来,可以提高抗骨折能力。远端基台螺丝。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号