首页> 外文期刊>Journal of Physics, D. Applied Physics: A Europhysics Journal >Assessment of S-phase coated medical grade stainless steel (Ortron 90) for use in the human joint replacement corrosion-wear environment
【24h】

Assessment of S-phase coated medical grade stainless steel (Ortron 90) for use in the human joint replacement corrosion-wear environment

机译:评估用于人体关节置换腐蚀磨损环境的S相涂层医用级不锈钢(Ortron 90)

获取原文
获取原文并翻译 | 示例

摘要

Hard and corrosion resistant coating materials such as S-phase, a nitrogen supersaturated Fe-Ni-Cr-Mo phase, have the potential to limit the corrosion-wear (tribo-corrosion) damage that can affect the metallic bearing surfaces of joint replacement devices in vivo, but hitherto have not been much investigated. To test their suitability, a range of S-phase coatings (with five nitrogen concentrations [N]) was applied to several polished Ortron 90 (biomedical grade austenitic stainless steel) test pieces via magnetron sputter deposition. Crevice attack, via intergranular corrosion, was found to be worse for most S-phase coated and uncoated Ortron 90 materials when tested in bovine serum compared with testing in 0.89% NaCl solution. Measurements of the breakdown potentials and the anodic passive current density, for a given test solution, revealed the crevice corrosion response of uncoated Ortron 90 to be more severe than the S-phase coated materials, although this was not unequivocally supported by 2D profilometry measurements. Corrosion-wear tests made using a sliding contact Al2O3 counterface showed the surface degradation to be more aggressive in saline solution than in bovine serum. All S-phase coatings were very effective in mitigating corrosion-wear of Ortron 90 in bovine serum solution, but only one S-phase coating composition, containing 21 at% N, gave improved resistance when tested in 0.89% NaCl solution.
机译:诸如S相(一种过饱和的Fe-Ni-Cr-Mo氮相)之类的坚硬且耐腐蚀的涂层材料有可能限制腐蚀磨损(摩擦腐蚀)的损害,这种损害会影响关节置换装置的金属轴承表面在体内,但迄今尚未进行大量研究。为了测试它们的适用性,通过磁控溅射沉积将一系列S相涂层(氮浓度为五个[N])施加到几个抛光的Ortron 90(生物医学级奥氏体不锈钢)试件上。与在0.89%NaCl溶液中进行测试相比,在牛血清中测试时,对于大多数S相涂层和未涂层​​Ortron 90材料,通过晶间腐蚀产生的缝隙腐蚀更为严重。对于给定的测试溶液,击穿电位和阳极无源电流密度的测量结果表明,未涂覆的Ortron 90的缝隙腐蚀响应比S相涂覆的材料更严重,尽管2D轮廓测量法并未明确支持。使用滑动接触式Al2O3相对表面进行的腐蚀磨损测试表明,盐溶液中的表面降解比牛血清中的腐蚀更剧烈。所有S相涂层在减轻Ortron 90在牛血清溶液中的腐蚀磨损方面都非常有效,但是当在0.89%NaCl溶液中测试时,只有一种含21 at%N的S相涂层组合物具有改进的电阻。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号