OBJECTIVE: The purpose of this study was to compare the accuracy and reliability of 2 well counter methods for measuring the activity concentration of (18)F-FDG in blood samples. METHODS: Three to 5 blood samples from 154 patient studies were weighed and measured in a well counter. The (18)F-FDG activity concentration was derived using, first, a direct calibration factor to convert measured well counter readings into activity concentration and, second, a comparison of measured counts with those of a specified standard solution. RESULTS: The ratio between the activity concentration results of the 2 methods was 0.996 +/- 0.033, indicating that the methods provided equal results. CONCLUSION: Because the standard solution method is more prone to human error, less reproducible, and more labor intensive, preference should be given to the direct calibration method.
机译:目的:本研究的目的是比较两种测定血样中(18)F-FDG活性浓度的井计数方法的准确性和可靠性。方法:对来自154个患者研究的3至5个血液样本进行称重并在井柜中进行测量。 (18)F-FDG活性浓度的计算方法如下:首先,使用直接校准因子将测得的孔计数器读数转换为活性浓度,其次,将测得的计数值与指定标准溶液的计数值进行比较。结果:两种方法的活性浓度结果之间的比率为0.996 +/- 0.033,表明这些方法提供了相等的结果。结论:由于标准溶液法更容易发生人为错误,重复性较低,劳动强度大,因此应优先选择直接校正方法。