New transcutaneous jaundice device with two optical paths.

机译:具有两个光路的新型经皮黄疸装置。

获取原文
获取原文并翻译 | 示例
           

摘要

OBJECTIVES: To confirm the accuracy and precision of transcutaneous bilirubin (TcB) values measured by a new device with two optical paths (JM-103) and the value of total serum bilirubin (TSB) level in clinical units of measurement. METHODS: For comparison of the levels of accuracy and precision of JM-103 and the old device (JM-102), serum samples were collected from 77 Japanese infants in three different hospitals including 24 preterm infants and 53 term infants. Measurement of TcB by JM-103 and JM-102 were performed on the forehead of each infant within 30 min before or after blood sampling. RESULTS: The range of TSB was limited to 19.6 mg/dL and to 17 mg/dL for preterm infants. The correlation coefficients for all subjects (r = 0.94) and for term and preterm subjects between TcB measured by JM-103 and TSB was higher than that between TcB measured by JM-102 and TSB. The regression line in term infants between TcB measured by JM-103 and TcB was similar to that in preterm infants. The error distribution of TcB measured by JM-102 and TSB for all subjects (0.00 +/- 2.21) and for term and preterm subjects was larger than that of TcB measured by JM-103 and TSB (all subjects, 0.30 +/- 1.55).
机译:目的:确认具有两个光路(JM-103)的新设备测量的经皮胆红素(TcB)值的准确性和准确性以及临床测量单位中血清总胆红素(TSB)水平的值。方法:为了比较JM-103和旧设备(JM-102)的准确性和精密度,从三所不同医院的77名日本婴儿中收集了血清样本,其中包括24名早产儿和53名足月儿。在采血之前或之后的30分钟内,对每个婴儿的前额进行JM-103和JM-102对TcB的测量。结果:早产儿的TSB范围被限制在19.6 mg / dL,17 mg / dL。 JM-103和TSB测得的TcB之间的所有受试者(r = 0.94)以及足月和早产受试者的相关系数均高于JM-102和TSB测得的TcB之间的相关系数。通过JM-103测量的TcB和TcB在足月婴儿中的回归线与早产婴儿中的回归线相似。通过JM-102和TSB测量的所有受试者(0.00 +/- 2.21)以及足月和早产受试者的TcB误差分布均大于通过JM-103和TSB测量的TcB的误差分布(所有受试者0.30 +/- 1.55 )。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号