Evaluation of the iCE280 Analyzer as a potential high-throughput tool for formulation development.

机译:评估iCE280分析仪作为潜在的高通量制剂开发工具。

获取原文
获取原文并翻译 | 示例
           

摘要

The iCE280 Analyzer (iCE280) was evaluated for its potential application as a high-throughput tool to determine pI and separate charge related species using glycosylated, non-glycosylated and pegylated protein therapeutics as models. Resolution was achieved for glycosylated and non-glycosylated molecules, but remained a challenge for pegylated proteins. The sources of charge variants were determined to be the presence of C-terminal lysine residues, sialic acid content, and deamidation. Limited assay performance evaluation demonstrated that the method was linear in the concentration range of 2-333 microg/ml of IgG with linear regression coefficients of 0.984, 0.998, and 0.990 for acidic, main and basic species, respectively. Limit of detection and limit of quantitation were determined to be 3 and 11 microg/ml. The R.S.D. for intra- and inter-day precision as well as reproducibility was determined to be 0.2% or less for all pI values and 1.4% or less for acidic and main peak area distribution; the R.S.D.for basic peak area distribution was 5.7% or less. Robustness testing was performed by deliberately deviating +/-50% of pharmalyte concentration away from the desired condition. This deviation revealed a pI shift of only 0.06 units and resulted in no significant impact on area percent distribution. Utilization of iCE280 Analyzer eliminated the mobilization step associated with traditional capillary isoelectric focusing analysis and increased analytical throughput at least 2-fold.
机译:使用糖基化,非糖基化和聚乙二醇化蛋白治疗剂作为模型,评估了iCE280分析仪(iCE280)作为高通量工具以确定pI和分离与电荷相关的物种的潜在应用。糖基化和非糖基化分子均实现了分离,但仍对聚乙二醇化蛋白构成了挑战。确定电荷变体的来源是C末端赖氨酸残基的存在,唾液酸含量和脱酰胺作用。有限的测定性能评估表明,该方法在2-333 microg / ml IgG浓度范围内是线性的,对于酸性,主要和碱性物质,线性回归系数分别为0.984、0.998和0.990。检测限和定量限确定为3和11 microg / ml。 R.S.D.对于日内和日间精度以及重现性,所有pI值均确定为0.2%或更低,而酸性峰和主峰面积分布的再现性被确定为1.4%或更低;基本峰面积的R.S.D.为5.7%以下。通过故意将+/- 50%的药剂浓度偏离所需的条件进行稳健性测试。该偏差表明pI偏移仅为0.06单位,并且对面积百分比分布没有显着影响。 iCE280分析仪的使用消除了与传统毛细管等电聚焦分析相关的移动步骤,并将分析通量提高了至少2倍。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号