Relationship between dental implant stability determined by resonance frequency analysis measurements and peri-implant vertical defects: an in vitro study.

机译:通过共振频率分析测量确定的种植体稳定性与种植体周围垂直缺陷之间的关系:一项体外研究。

获取原文
获取原文并翻译 | 示例
           

摘要

Resonance frequency (RF) analysis is a non-invasive and objective technique developed for implantology, where it measures the stability of the implant in the bone socket. A limited number of studies were performed to compare different stability measurement devices. The aim of the study was to evaluate the association between the implant stability and peri-implant vertical defect by using the new wireless RF analyser. Micro-textured rough-surfaced implants with diameters of 3.7 and 4.8 mm and with a length of 12 mm were used. The peak insertion torque was recorded with the help of the torque control system only during the implant placement. The peri-implant defects were created in millimetre increments ranging between 0 and 5 mm to the same extent on all implants. The RF analysis values were measured by using the RF analyser. Significant correlation was found between insertion torque (44.3 +/- 3 N cm) and implant stability quotient (ISQ) (all implants: r = 0.76, 3.7 mm diameter: r = 0.65, 4.8 mm diameter: r = 0.80). For 3.7 x 12-mm(2) implants, the mean ISQ values were 72.6 +/- 2.4, 69.8 +/- 2, 67.2 +/- 1.8, 64.3 +/- 1.9, 61.2 +/- 2 and 57.2 +/- 2.5 when peri-implant vertical defects were 0, 1, 2, 3, 4 and 5 mm; and the corresponding values were 76.5 +/- 2.2, 74.5 +/- 2.2, 72.3 +/- 2, 70 +/- 1.9, 67 +/- 1.9 and 63.4 +/- 2.3, respectively, for 4.8 x 12-mm(2) implants. All values were significantly different when compared with each other. The wireless RF analyser seems to be a suitable and reliable device to determine the implant stability. Peri-implant bone loss simulated by using acrylic models may result with a decrease in ISQ values for osseo-integrated implants measured by the RF analyser.
机译:共振频率(RF)分析是为植入术开发的一种非侵入性客观方法,用于测量植入物在骨槽中的稳定性。进行了有限数量的研究以比较不同的稳定性测量设备。该研究的目的是通过使用新型无线射频分析仪评估植入物稳定性与植入物周围垂直缺损之间的关系。使用直径为3.7和4.8毫米,长度为12毫米的微纹理粗糙表面植入物。仅在植入物放置期间,借助于扭矩控制系统记录峰值插入扭矩。在所有植入物上,植入物周围的缺损均以相同的程度以0到5 mm的毫米增量产生。通过使用RF分析仪测量RF分析值。发现插入扭矩(44.3 +/- 3 N cm)与植入物稳定性商(ISQ)之间存在显着相关性(所有植入物:r = 0.76,直径3.7 mm:r = 0.65,直径4.8 mm:r = 0.80)。对于3.7 x 12毫米(2)植入物,平均ISQ值为72.6 +/- 2.4、69.8 +/- 2、67.2 +/- 1.8、64.3 +/- 1.9、61.2 +/- 2和57.2 +/- 2.5当种植体周围垂直缺陷为0、1、2、3、4和5 mm时;对于4.8 x 12毫米(),相应的值分别为76.5 +/- 2.2、74.5 +/- 2.2、72.3 +/- 2、70 +/- 1.9、67 +/- 1.9和63.4 +/- 2.3 2)植入物。相互比较时,所有值均显着不同。无线射频分析仪似乎是确定植入物稳定性的合适且可靠的设备。使用丙烯酸模型模拟的种植体周围骨质损失可能导致RF分析仪测量的骨整合种植体的ISQ值降低。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号