Stability of volar locking plate systems for AO type C3 fractures of the distal radius: biomechanical study in a cadaveric model.

机译:vol骨锁定板系统对AO型C3远端radius骨骨折的稳定性:尸体模型中的生物力学研究。

获取原文
获取原文并翻译 | 示例
           

摘要

BACKGROUND: The purpose of the present study was to compare the relative stability of five volar locking plates (all of which are available for the treatment of intraarticular fractures of the distal radius) under loading conditions simulating the physiological forces that occur during early active mobilization. METHODS: Five plating techniques were applied to surgically simulated AO type C3.2 distal radius fractures in formalin-fixed cadavers. The specimens were tested with a servohydraulic materials testing machine with 250 N of axial compression load for 3000 cycles. After cyclic loading, the specimens were loaded until they demonstrated failure in axial compression. The five fixation systems studied included a DRV locking plate (group 1), a Stellar plate (group 2), an Acu-Loc plate (group 3), AO Locking Distal Radius System 2.4 (group 4); and a Matrix SmartLock plate (group 5). RESULTS: None of the plate fixations tested failed during the cyclic loading. Group 2 had a higher elastic limit than groups 4 and 5. There were no significant differences among the five groups for the failure load. Failure occurred at the distal portion of the fixation system, at the ulnar side locking pin, or the locking screw was bent (groups 1, 2, 3); the ulnar side locking screw was broken (groups 3, 4); the locking screw became loose (group 4); and the ulnar side locking screw was uncoupled from the screw hole (group 5). CONCLUSIONS: All of the five volar plate fixation systems provided sufficient stability to permit 3000 repeated motions of the digits after surgery for AO type C3 distal radius fractures.
机译:背景:本研究的目的是在模拟早期主动动员过程中产生的生理力的载荷条件下,比较五个掌侧锁定板(所有这些均可用于治疗radius骨远端关节内骨折)的相对稳定性。方法:采用五种电镀技术,对福尔马林固定尸体中的手术模拟的AO C3.2型distal骨远端骨折进行了手术。用伺服液压材料试验机在250 N的轴向压缩载荷下对样品进行3000个循环的测试。循环加载后,加载标本,直到显示轴向压缩失败为止。研究的五个固定系统包括DRV锁定板(组1),Stellar板(组2),Acu-Loc板(组3),AO锁定远端半径系统2.4(组4);和Matrix SmartLock板(组5)。结果:在循环加载过程中,测试的板固定均未失败。第2组的弹性极限高于第4组和第5组。五组之间的破坏载荷无显着差异。在固定系统的远端,尺侧锁定销处发生故障,或者锁定螺钉弯曲(1、2、3组);尺侧锁定螺钉损坏(第3、4组);锁紧螺丝松动(组4);并从螺钉孔(第5组)上松开尺侧锁定螺钉。结论:五个掌侧钢板固定系统均提供了足够的稳定性,以允许在AO型C3 radius骨远端骨折手术后进行3000次手指重复运动。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号