Intravascular ultrasound Doppler measurement of blood flow velocity.

机译:血管内超声多普勒测量血流速度。

获取原文
获取原文并翻译 | 示例
           

摘要

BACKGROUND AND PURPOSE: The authors investigated the feasibility and accuracy of intravascular Doppler sonography (IVDS) with a newly developed microprobe. METHODS: The known method to determine blood flow velocity by transcutaneous Doppler sonography was transferred to an intravascular usable Doppler probe. With the improved technique, a microprobe measuring 0.3 mm in diameter can be advanced through a 5F catheter used in routine diagnostic angiography to perform intravascular Doppler sonography. In a first step, the conditions for application and measurement were studied in a flow tube model and were transferred in a second step to patients undergoing routine angiography and patients with arterial stenoses. RESULTS: Measurements with nondegassed liquids showed a high intensive acoustic signal, but no echo effect could be recorded in degassed liquids. IVDS of healthy vessels in 40 patients showed the same typical flow pulse curve as seen in transcutaneous measurements. The optimal position of the microprobe with respect to the contact of the vessel wall, the diastolic and systolic phase, and the distance to the tip of the catheter could be evaluated. In 95.6% of the 40 patients with healthy vessels, IVDS was successful. Fourteen patients with arterial stenotic diseases were investigated before and after percutaneous transluminal angioplasty, and the results of these investigations correlated well with the angiographical results. CONCLUSIONS: For percutaneously directly reachable vessels, the transcutaneous Doppler sonography is the choice for easy noninvasive and inexpensive measurement of blood flow velocity. However, for vessels that are difficult or impossible to reach percutaneously, intravascular measurement is a valid procedure.
机译:背景与目的:作者研究了使用最新开发的微探针进行血管内多普勒超声检查(IVDS)的可行性和准确性。方法:将经皮多普勒超声确定血流速度的已知方法转移到血管内可用的多普勒探针中。利用改进的技术,可以通过常规诊断性血管造影术中使用的5F导管推进直径为0.3 mm的微探针,以进行血管内多普勒超声检查。第一步,在流量管模型中研究应用条件和测量条件,并在第二步中将其转移至接受常规血管造影术的患者和患有动脉狭窄的患者。结果:使用非脱气液体进行的测量显示出高强度的声信号,但在脱气液体中未记录回声效应。 40例患者的健康血管的IVDS显示出与经皮测量相同的典型流量脉搏曲线。可以评估微探针相对于血管壁,舒张期和收缩期的最佳位置,以及与导管尖端的距离。在40例健康血管患者中,有95.6%的患者IVDS成功。在行经皮腔内血管成形术之前和之后对14例患有动脉狭窄疾病的患者进行了研究,这些研究的结果与血管造影结果密切相关。结论:对于经皮直接可达血管,经皮多普勒超声检查是简便,无创且廉价地测量血流速度的选择。但是,对于难以或无法通过皮肤到达的血管,血管内测量是有效的方法。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号