A comparison of reflectance pulse oximetry and near infrared spectroscopy for the determination of the fraction of oxygenated haemoglobin at low oxygen saturations in an animal study

机译:在动物研究中,比较反射式脉搏血氧饱和度法和近红外光谱法测定低血氧饱和度下氧合血红蛋白的比例

获取原文
获取原文并翻译 | 示例
           

摘要

Near infrared spectroscopy scans of cerebral tissue in vivo were compared to reflectance pulse oximetry of local tissue by comparing each method to in vitro femoral arterial blood measurements in an experimental study on 16 newborn piglets. To test the accuracy, the 16 piglets were desaturated step wise during general anaesthesia, leading to the following results of linear regression analysis of complete data, linear regression analysis for individual piglets and bias ± precision (1SD). Local oxygen saturation (reflectance pulse oximetry) versus arterial (haemoximeter): linear regression analysis R{sup}2 = 0.92, n = 207, linear regression analysis for individual piglets: R{sup}2 = 0.95±0.05{sup}*, p < 0.05 versus near infrared spectroscopy, bias ± precision (1SD): 0.15±9.65%; cerebral tissue oxygenation index (near infrared spectroscopy) versus arterial (haemoximeter): linear regression analysis R{sup}2 = 0.26, n = 161, linear regression analysis for individuals R{sup}2 = 0.83 ±0.15, bias ± precision (1SD) -3.33 ±29%; cerebral concentration of delta oxygenated haemoglobin (near infrared spectroscopy, NIRO 300) versus the corresponding arterial parameter (haemoximeter): linear regression analysis R{sup}2 = 0.76, n = 189, linear regression analysis for individuals R{sup}2 = 0.90 ±0.06, bias ± precision (1SD) -2.54±3μMcm{sup}(-3). The results of reflectance pulse oximetry in local tissue correlated more closely to in vitro femoral arterial blood measurements (haemoximeter) than results of near infrared spectroscopy, in vivo.
机译:在16个新生仔猪的实验研究中,通过将每种方法与体外股动脉血测量值进行比较,将体内脑组织的近红外光谱扫描与局部组织的反射脉冲血氧饱和度进行了比较。为了测试准确性,对16头仔猪在全身麻醉期间逐步进行去饱和,得出以下完整数据的线性回归分析,单个仔猪的线性回归分析和偏差±精度(1SD)的结果。局部血氧饱和度(反射脉冲血氧饱和度)与动脉血氧饱和度(血氧仪):线性回归分析R {sup} 2 = 0.92,n = 207,个别仔猪的线性回归分析:R {sup} 2 = 0.95±0.05 {sup} *,与近红外光谱相比,p <0.05,偏差±精度(1SD):0.15±9.65%;脑组织氧合指数(近红外光谱)与动脉(血氧计):线性回归分析R {sup} 2 = 0.26,n = 161,个体的线性回归分析R {sup} 2 = 0.83±0.15,偏差±精度(1SD )-3.33±29%;氧合血红蛋白的大脑浓度(近红外光谱法,NIRO 300)与相应的动脉参数(血氧计):线性回归分析R {sup} 2 = 0.76,n = 189,个体的线性回归分析R {sup} 2 = 0.90 ±0.06,偏置±精度(1SD)-2.54±3μMcm{sup}(-3)。与体内近红外光谱法的结果相比,局部组织中反射脉冲血氧仪的结果与体外股动脉血测量(血氧计)的相关性更高。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号