In a microbiological sample study of 15 wet-primed cardiopulmonary bypass circuits in standby mode for 72 hours under regular clinical conditions, no contamination of the priming fluid or the connectors could be detected. Hand contact surfaces of the machines demonstrated environmental microorganisms. These findings indicate the safe use of primed cardiopulmonary bypass circuits in standby mode for 72 hours. A surface disinfection of hand contact surfaces immediately before use is recommended.
机译:在常规临床条件下,在待机模式下对15个湿式启动式心肺旁路回路进行了72小时的微生物学样本研究,未检测到启动液或连接器受到污染。机器的手接触表面显示出环境微生物。这些发现表明在待机模式下安全使用带预备心肺旁路电路72小时。建议在使用前立即对手接触表面进行表面消毒。