Proton spectroscopic imaging metabolite ratios in human brain: measurement error proportional to the mean.

机译:人脑中质子光谱成像代谢物的比率:与平均值成正比的测量误差。

获取原文
获取原文并翻译 | 示例
           

摘要

As part of our 3D proton MR spectroscopic imaging (MRSI) studies of intracranial tumors, we carried out a repeatability study in normal volunteers. While analyzing these data we detected an interesting statistical observation: measurement error proportional to the mean, which, to the best of our knowledge, has not been documented before and which may, therefore, be of interest. While we acknowledge that this observation might be specific to our particular combination of hardware and pulse sequence parameters, we feel that the result is still noteworthy.
机译:作为我们对颅内肿瘤的3D质子MR光谱成像(MRSI)研究的一部分,我们在正常志愿者中进行了可重复性研究。在分析这些数据时,我们发现了一个有趣的统计观察结果:与平均值成正比的测量误差,据我们所知,该误差以前没有得到记录,因此可能引起关注。尽管我们承认该观察可能是特定于我们的硬件和脉冲序列参数的特定组合,但我们认为结果仍然值得关注。