首页> 外文期刊>Journal of magnetic resonance imaging: JMRI >Quantitative assessment of cardiac output and left ventricular function by noninvasive phase-contrast and cine MRI: validation study with invasive pressure-volume loop analysis in a swine model.
【24h】

Quantitative assessment of cardiac output and left ventricular function by noninvasive phase-contrast and cine MRI: validation study with invasive pressure-volume loop analysis in a swine model.

机译:通过无创相衬和电影MRI定量评估心输出量和左心室功能:猪模型中有创压力容积环分析的验证研究。

获取原文
获取原文并翻译 | 示例
           

摘要

PURPOSE: To validate noninvasive cardiac output measurements of phase-contrast magnetic resonance imaging (PC-MRI) and cine MRI using an invasive pressure-volume (PV) loop technique on a swine model. MATERIALS AND METHODS: We compared three methods for evaluating cardiac function at rest and under pharmaceutical low-dose inotropic infusion conditions: 1) phase-contrast MRI, 2) cine MRI, and 3) PV loop relationship. These measurements were made in 14 domestic pigs under rest conditions. Identical MRI acquisitions and PV loop analysis were performed on six pigs from the same group that received an infusion of dobutamine 2.5 mug/kg/min. Cardiac outputs from all measurements were analyzed and compared using linear regression and Bland-Altman analysis. RESULTS: Noninvasive PC-MRI and cine MRI did not show any significant differences compared to an invasive PV loop technique for measurement of cardiac output under both rest (PC-MRI, cine MRI, and PV loop, 3.17 +/- 0.45, 3.18 +/- 0.61, 3.45 +/- 0.41 L/min, respectively) and pharmaceutical low-dose inotropic infusion conditions (PC-MRI, cine MRI, and PV loop, 4.78 +/- 0.53, 4.7 +/- 0.6, 4.96 +/- 0.48 L/min, respectively). Statistical analysis showed good agreement of cardiac output measurements at rest (R(2) = 0.83) and under low-dose inotropic infusion conditions (R(2) = 0.74) using PC-MRI and PV loop techniques. Cardiac output measurement using cine MRI and PV loop techniques also showed good agreement at rest (R(2) = 0.85) and under low-dose inotropic infusion conditions (R(2) = 0.76). Furthermore, cardiac outputs determined with the three modalities showed good agreement over a wide range of heart rates (90-180 bpm). CONCLUSION: MRI can provide a reliable, noninvasive measurement of cardiac output that can be carried out without the complications that are inherent with current invasive procedures.
机译:目的:为了验证在猪模型上使用有创压力体积(PV)循环技术的相衬磁共振成像(PC-MRI)和电影MRI的无创心输出量测量。材料与方法:我们比较了三种评估静息状态和药物低剂量正性输注条件下心功能的方法:1)相差MRI,2)电影MRI和3)PV回路关系。这些测量是在休息条件下对14头家猪进行的。对同一组的六只猪进行了相同的MRI采集和PV环路分析,这些猪接受了2.5杯/千克/分钟的多巴酚丁胺输注。使用线性回归和Bland-Altman分析对所有测量的心脏输出进行分析和比较。结果:与侵入式PV环技术相比,无创PC-MRI和电影MRI在两种休息状态下的心输出量测量均无明显差异(PC-MRI,电影MRI和PV环,3.17 +/- 0.45,3.18 + /-分别为0.61、3.45 +/- 0.41 L / min)和药物低剂量正性输注条件(PC-MRI,电影MRI和PV回路),4.78 +/- 0.53,4.7 +/- 0.6,4.96 + / -分别为0.48 L / min)。统计分析表明,使用PC-MRI和PV回路技术,在静息时(R(2)= 0.83)和低剂量正性输注条件下(R(2)= 0.74)的心输出量测量结果具有良好的一致性。使用电影MRI和PV回路技术进行的心输出量测量也显示出在静息状态(R(2)= 0.85)和低剂量正性输注条件下(R(2)= 0.76)的良好一致性。此外,用这三种方式确定的心输出量在广泛的心率范围内(90-180 bpm)显示出良好的一致性。结论:MRI可以提供可靠的,无创的心输出量测量结果,而无需进行当前有创程序固有的并发症。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号