Local flip angle correction for improved volume T1-quantification in three-dimensional dGEMRIC using the Look-Locker technique.

机译:使用Look-Locker技术改进三维dGEMRIC中体积T1量化的局部翻转角校正。

获取原文
获取原文并翻译 | 示例
           

摘要

PURPOSE: To present an evaluation method for three-dimensional Look-Locker (3D-LL) based T1 quantification, calculating correct T1 values independent of local flip angle (FA) variations. The method was evaluated both in phantoms and in vivo in a delayed Gadolinium Enhanced MRI of Cartilage (dGEMRIC) study with 33 subjects. MATERIALS AND METHODS: T1 was measured with 3D-LL, using both local FA correction and a precalculated FA slice profile, and compared with standard constant FA correction, for all slices in phantoms and in both femur condyles in vivo. T1 measured using two-dimensional Inversion Recovery (2D-IR) was used as gold standard. RESULTS: Due to the FA being slice dependent, the standard constant FA correction results in erroneous T1 (systematic error = 109.1 ms in vivo), especially in the outer slices. With local FA correction, the calculated T1 is excellent for all slices in phantoms (<5% deviation from 2D-IR). In vivo the performance is lower (systematic error = -57.5 ms), probably due to imperfect inversion. With precalculated FA correction the performance is very good also in vivo (systematic error = 13.3 ms). CONCLUSION: With the precalculated FA correction method, the 3D-LL sequence is robust enough for in vivo dGEMRIC, even outside the centermost slices.
机译:目的:提出一种基于三维Look-Locker(3D-LL)的T1量化的评估方法,计算独立于局部翻转角(FA)变化的正确T1值。该方法已在33位受试者的延迟d增强MRI软骨研究(dGEMRIC)中在幻像和体内进行了评估。材料与方法:使用3D-LL测量T1,同时使用局部FA校正和预先计算的FA切片轮廓,并与标准恒定FA校正相比较,用于体模中和体内两个股骨dy中的所有切片。使用二维反转恢复(2D-IR)测量的T1用作金标准。结果:由于FA依赖于切片,因此标准的恒定FA校正会导致错误的T1(体内系统误差= 109.1 ms),尤其是在外部切片中。通过局部FA校正,对于幻像中的所有切片,计算出的T1都非常出色(与2D-IR的偏差小于5%)。在体内,性能可能较低(系统误差= -57.5 ms),这可能是由于不完全反转所致。通过预先计算的FA校正,在体内的性能也非常好(系统误差= 13.3 ms)。结论:通过预先计算的FA校正方法,即使在最中心的切片之外,3D-LL序列对于体内dGEMRIC也足够鲁棒。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号