This investigation evaluated metallic surgical instruments for magnetic resonance (MR) safety in association with a 1.5-Tesla/64-MHz MR system. Seven different instruments (mallet, bone punch, curette, Weil-Blakesley ethmoid forceps, suction cannula, septum speculum, and Kocher-Langenbeck retractor; Aesculap, Inc. (South San Francisco, CA) were tested for magnetic field interactions, heating, and generation of artifacts by using previously described techniques. Heating was evaluated for the septum speculum and Kocher-Langenbeck retractor by using a special gel-filled phantom and a fluoroptic thermometer to record temperatures immediately before and during MRI performed at a whole-body averaged SAR of 1.3 W/kg. Artifacts were assessed with the instruments placed inside of a gel-filled phantom and performing MRI using T1-weighted spin-echo and gradient-echo pulse sequences. Magnetic field interactions were relatively minor (deflection angles, 0 to 7 degrees; torque, 0 to +1), the highest temperature changes were < or = +0.8 degrees C, and the artifacts should not create substantial problems considering the "intended use" for these instruments. The findings of the MR safety tests indicated that the seven different metallic surgical instruments (Aesculap, Inc.) would be safe and acceptable for use in interventional MRI procedures performed with MR systems with static magnetic fields of 1.5 T or less. J. Magn. Reson. Imaging 2001;13:152-157. Copyright 2001 Wiley-Liss, Inc.
机译:这项研究评估了金属手术器械与1.5-Tesla / 64-MHz MR系统的磁共振(MR)安全性。测试了七种不同的仪器(麦芽,骨头打孔器,刮匙,Weil-Blakesley筛骨钳,吸管,隔膜隔膜和Kocher-Langenbeck牵开器; Aesculap,Inc.(加利福尼亚州南旧金山)的磁场相互作用,加热和通过使用特殊的凝胶填充体模和荧光温度计测量隔垫窥镜和Kocher-Langenbeck牵开器的加热,以记录在MRI之前和期间以全身平均SAR记录的温度。 1.3 W / kg。使用置于凝胶填充的体模内的仪器评估伪像,并使用T1加权自旋回波和梯度回波脉冲序列进行MRI,磁场相互作用相对较小(偏转角为0至7度) ;扭矩,从0到+1),最高温度变化为<或= +0.8摄氏度,考虑到“预期用途”,工件不应造成实质性问题。 se仪器。 MR安全性测试的结果表明,七种不同的金属手术器械(Aesculap,Inc.)对于在静态磁场为1.5 T或更小的MR系统执行的介入性MRI程序中将是安全且可接受的。 J.Magn。雷森成像2001; 13:152-157。版权所有2001 Wiley-Liss,Inc.

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号