PURPOSE: To validate the accuracy of T1 measurement by three-dimensional Look-Locker method (3D LL) for delayed gadolinium-enhanced MRI of cartilage (dGEMRIC) of human subjects with and without osteoarthritis (OA), as compared with two-dimensional inversion recovery fast spin-echo (2D IR-FSE) technique. MATERIALS AND METHODS: MR sagittal images of the knees were acquired for T1 mapping in 29 subjects with standard 2D IR-FSE and 3D LL sequences 90-135 min following administration of 0.2 mmol/kg Gd-DTPA(2-). T1 maps of femoral and tibial cartilage were generated using custom software. Comparisons in T1 values between the two techniques were performed using regression analysis. RESULTS: Good agreement in T1 values between 2D IR-FSE and 3D LL was observed (R values of 0.90, and 0.85, and 0.86 for all, OA, and control subjects, respectively) when acquired within 15 min. CONCLUSION: The 3D LL sequence provides accurate T1 estimates of articular cartilage with advantages of entire joint coverage, shorter acquisition time, and a wide range of inversion times sampled within a single acquisition. J. Magn. Reson. Imaging 2008. (c) 2008 Wiley-Liss, Inc.
机译:目的:通过二维Look-Locker方法(3D LL)验证带有或不带有骨关节炎(OA)的人类受试者软骨延迟delayed增强MRI(dGEMRIC)与二维倒置相比的准确性恢复快速自旋回波(2D IR-FSE)技术。材料与方法:在接受0.2 mmol / kg Gd-DTPA(2-)后90-135分钟内,以标准2D IR-FSE和3D LL序列在29位受试者中获取T1作图的膝盖MR矢状面图像。使用定制软件生成了股骨和胫骨软骨的T1图。两种技术之间的T1值比较使用回归分析进行。结果:在15分钟内获得2D IR-FSE和3D LL之间的T1值时,有很好的一致性(R值分别为OA和对照组的R值分别为0.90、0.85和0.86)。结论:3D LL序列提供了准确的关节软骨T1估计值,具有整个关节覆盖,更短的采集时间以及在一次采集中采样范围广泛的反转时间的优势。 J.Magn。雷森影像杂志2008。(c)2008 Wiley-Liss,Inc.。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号