Cyclic loads do not compromise functionality of the interspinous spacer or cause damage to the spinal segment: An in vitro analysis

机译:循环载荷不会损害棘突间间隔物的功能或不会损害脊柱节段:体外分析

获取原文
获取原文并翻译 | 示例
           

摘要

Study Design. A biomechanical study to evaluate the effects of interspinous spacer under cyclic complex loading. Objectives. To determine the risk of device migration and to assess damage on the device and specimen under extreme coupled motion. Another objective was to evaluate the effect on vertebral foramen and canal dimensions after spacer implantation. Summary of Background Data. Interspinous spacers are a relatively new treatment option that are clinically exposed to complex loads. However, the biomechanical performance of these spacers has not been well characterized. Methods. Six human cadaveric motion segments were used for this study. The interspinous spacer (SuperionTM, Vertiflex Inc, California) was tested for 5°extension/10° flexion coupled with an axial rotation of 3°. CT images were taken for specimens in neutral, 5° extension, and 10° flexion before and after the implantation of the spacer. Vertebral foramen and canal dimensions were quantified. Results. There was no device migration or subsidence. Specimens did not sustain any significant injury during testing. Canal area was minimally altered and foramen height, width, and area increased in extension and were statistically significant as compared to intact. Conclusion. Interspinous spacer effectively prevents the motion at the implanted level and does not change the anatomy significantly.
机译:学习规划。评估棘突间间隔物在循环复合物载荷作用下的生物力学研究。目标。确定设备迁移的风险并评估极端耦合运动下设备和样品的损坏。另一个目的是评估间隔物植入后对椎孔和管径的影响。背景数据摘要。棘突间隔物是一种相对较新的治疗选择,在临床上会暴露于复杂的负荷下。但是,这些间隔物的生物力学性能尚未很好地表征。方法。六个人体尸体运动段用于这项研究。对棘突间垫片(SuperionTM,Vertiflex Inc,加利福尼亚)进行了5°延伸/ 10°弯曲和3°轴向旋转的测试。在植入垫片之前和之后,以中性,5°延伸和10°屈曲的标本拍摄CT图像。量化椎孔和管的尺寸。结果。没有设备迁移或下沉。标本在测试过程中没有遭受任何重大伤害。与完整相比,根管面积变化最小,并且孔的高度,宽度和面积在延伸处增加,并且具有统计学意义。结论。棘突间垫片有效地防止了植入水平的运动,并且不会显着改变解剖结构。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号