The purpose of this study was to mechanically test a novel Unthreaded Expandable Fastener (UEF), manufactured using Selective Laser Melting, which was designed for fixation in the cervical lateral mass. The pull-out strength and stiffness of the prototype UEFs was evaluated in a non-osteoporotic ovine bone model against equivalent screws. The prototype UEF demonstrated a 41% increase in failure force and a 60% reduction in failure force standard deviation compared to the screws. All bone samples were micro CT-scanned and no significant differences in bone microstructural properties was found between the screw and UEF sample sets, indicating that the UEFs may be less sensitive to bone quality variation. This increased performance can potentially translate into improved surgical outcome and reduced surgical risk for lateral mass fixation. With further design optimisation, additional improvement in performance over screws may be possible in future studies.
机译:这项研究的目的是对使用选择性激光熔化技术制造的新型无螺纹可膨胀紧固件(UEF)进行机械测试,该紧固件旨在固定在颈椎侧块中。在非骨质疏松性羊骨模型中使用等效螺钉评估了原型UEF的拉出强度和刚度。与螺钉相比,原型UEF的破坏力提高了41%,破坏力标准偏差降低了60%。所有骨样品均进行了微CT扫描,在螺钉和UEF样品组之间未发现骨微结构特性的显着差异,这表明UEF对骨质量变化的敏感性较低。这种提高的性能可能潜在地转化为手术效果的改善,并降低了侧块固定的手术风险。随着进一步的设计优化,在未来的研究中可能会比螺钉的性能进一步提高。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号