An unusual resolution of T-wave oversensing in an implantable cardioverter defibrillator in a child with long QT syndrome.

机译:患有长QT综合征的儿童的植入式心脏复律除颤器中T波感应异常的解决方案。

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

T-wave oversensing poses a clinical challenge often resulting in inappropriate therapies and shocks. T-wave oversensing can often be resolved by certain programmable adjustments while accepting a small risk of undersensing ventricular fibrillation. The following case depicts a girl with LQT whose transvenous defibrillator repeatedly demonstrated T-wave oversensing despite optimizing various programmable features and a separate pace/sense lead. Utilization of an LV epicardial pace/sense lead with stable R-waves affirmed an unusual resolution of T-wave oversensing and avoidance of inappropriate shocks.
机译:T波感应过度带来临床挑战,通常会导致不适当的疗法和电击。 T波过度感测通常可以通过某些可编程的调整来解决,同时又可以接受较小的心室颤动感知风险。以下案例描述了一个患有LQT的女孩,尽管优化了各种可编程功能和单独的步速/感觉导联,但其静脉除纤颤器反复显示出T波感应过度。使用具有稳定R波的LV心外膜起搏/感觉导联可以肯定T波过度感应的异常分辨率并避免了不适当的电击。