Sphygmomanometers are frequently contaminated with bacteria and are implicated in the transmission of microbes. A pilot study was conducted to determine the effectiveness and feasibility of using a physical barrier device in reducing rates of microbial contamination of sphygmomanometer cuffs. Blood pressure recordings were made with and without a disposable plastic device on healthy individuals to ensure that it did not interfere with the accuracy of readings. Baseline rates of microbial colonisation of sphygmomanometer cuffs were obtained over a six week period. For six weeks the device was applied to patients' arms before measuring blood pressure, and sphygmomanometer cuffs were recultured during the intervention period. No difference was found in blood pressure recordings with and without the device. No difference in rates of bacterial contamination was observed during the intervention phase of the study. A disposable plastic device acting as a physical barrier may not reduce surface contamination of sphygmomanometer cuffs.
机译:血压计经常被细菌污染,并与微生物的传播有关。进行了一项初步研究,以确定使用物理屏障设备降低血压计袖带微生物污染率的有效性和可行性。在有或没有一次性塑料装置的情况下,对健康个体进行血压记录,以确保其不会干扰读数的准确性。在六周的时间内获得了血压计袖带的微生物定植基线速率。在测量血压之前,先将设备戴在患者手臂上六周,然后在干预期间对血压计袖带进行再培养。使用和不使用该设备的血压记录均未发现差异。在研究的干预阶段,未观察到细菌污染率的差异。用作物理屏障的一次性塑料装置可能无法减少血压计袖带的表面污染。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号