Interfragmentary compression across a simulated scaphoid fracture--analysis of 3 screws.

机译:模拟舟骨骨折的片段间压迫-分析3个螺钉。

获取原文
获取原文并翻译 | 示例
           

摘要

PURPOSE: To measure the interfragmentary compression generated across a simulated fracture in cadaveric scaphoids by 3 different headless compression screws. METHODS: A transverse osteotomy was made through the waist of each scaphoid and a load cell to measure compression was interposed between the fragments, which were then fixed internally retrograde with either an Acutrak Standard (n = 10), Acutrak Mini (n = 12), or Bold (n = 10) screw. The surgeon was blinded to the measured compression, which was recorded during screw insertion and for the following 5 minutes. As a measure of scaphoid bone quality the modulus of elasticity of the trabecular bone of each specimen was then calculated from uniaxial compression tests. RESULTS: The mean interfragmentary compression generated by the Acutrak Standard screw was significantly greater and more consistent than the Bold screw or the Acutrak Mini screw. The compression achieved by the Acutrak Standard screw was also more consistent than that obtained by eitherthe Bold or the Acutrak Mini screws as reflected by the lower standard deviation. The mean modulus of elasticity of the scaphoid trabecular bone was similar for each screw group. CONCLUSIONS: The interfragmentary compression generated by the Acutrak Standard screw was significantly greater and more consistent than that generated by either the Bold or Acutrak Mini screws. The compression generated by the Acutrak Standard and Mini screws was significantly better sustained over time than that generated by the Bold screw; however, these differences were small and may not be clinically important. The optimal compression required to promote scaphoid fracture union remains unknown and clinical trials are needed to further evaluate the outcome from using these devices.
机译:目的:通过3个不同的无头压缩螺钉来测量尸骨舟骨中模拟骨折产生的碎片间压缩。方法:在每个舟骨的腰部进行横向截骨术,并在各碎片之间插入一个测压元件以测量压缩力,然后用Acutrak Standard(n = 10),Acutrak Mini(n = 12)进行内部逆行固定。 ,或加粗(n = 10)螺钉。外科医生对所测得的压缩力不知情,该压缩力在螺钉插入过程中以及随后的5分钟内记录下来。然后,根据单轴压缩试验计算出每个标本的小梁骨的弹性模量,作为舟骨骨质量的量度。结果:Acutrak Standard螺钉产生的平均股间压缩比Bold螺钉或Acutrak Mini螺钉大得多,并且更加一致。 Acutrak Standard螺钉获得的压缩也比Bold或Acutrak Mini螺钉获得的压缩更一致,这反映出较低的标准偏差。每个螺钉组的舟骨小梁骨的平均弹性模量相似。结论:与粗体或Acutrak Mini螺钉相比,Acutrak Standard螺钉产生的片段间压缩明显更大且更一致。随着时间的推移,Acutrak Standard和Mini螺钉产生的压缩比Bold螺钉产生的压缩明显更好。但是,这些差异很小,在临床上可能并不重要。促进舟骨骨折愈合所需的最佳压缩力仍然未知,需要进行临床试验以进一步评估使用这些装置的结果。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号