Volar plating of AO C3 distal radius fractures: biomechanical evaluation of locking screw and locking smooth peg configurations.

机译:AO C3 radius骨远端骨折的掌侧钢板:锁定螺钉和锁定光滑钉结构的生物力学评估。

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

PURPOSE: The goal of this study was to determine whether locking screws or smooth locking pegs optimize fixation of AO C3 intra-articular distal radius fractures. A secondary goal was to determine which combinations of locking screws and smooth locking pegs influence construct stability. METHODS: In anatomic radius models, AO C3 intra-articular distal radius fractures were fixed using volar locking plates. For the first part, 16 specimens were randomized to receive either 2 locking screws or 2 smooth locking pegs in each of the 3 pairs of holes in the plate. For the second part, 30 specimens were randomized to receive any 4 combinations of locking screws and smooth locking pegs in each of the 3 pairs of holes. Axial loading to failure was applied. RESULTS: Constructs consisting of 4 smooth locking pegs within the lunate fragment were significantly weaker than constructs with 4 locking screws (means 626 N vs 981 N, respectively). Constructs with smooth locking pegs in the ulnar positions of the lunate fragment were weaker than with locking screws in these positions (means 737 N vs 977 N, respectively). Locking screws in the subchondral position of the lunate fragment were stronger than smooth locking pegs in these positions (means 1,227 N vs 934 N, respectively) and any other combination (means 1,227 N vs 942 N, respectively). CONCLUSIONS: Use of locking screws as opposed to smooth locking pegs for AO C3 intra-articular distal radius fractures, particularly subchondral and in the ulnar side of the lunate fragment, optimizes construct stability. This may have implications on postoperative rehabilitation protocols and may limit costs related to use of volar locking plates.
机译:目的:本研究的目的是确定锁定螺钉或光滑锁定钉是否优化了AO C3关节内distal骨远端骨折的固定。第二个目标是确定哪种锁紧螺钉和光滑的锁紧销组合会影响结构的稳定性。方法:在解剖radius骨模型中,使用掌侧锁定板固定AO C3关节内distal骨远端骨折。对于第一部分,将16个样本随机分配在板上的3对孔中的每对中分别接受2个锁定螺钉或2个平滑锁定钉。对于第二部分,将30个样本随机分配,以在3对孔中的每对中接受锁定螺钉和光滑锁定钉的任意4种组合。施加轴向载荷直至失效。结果:在月牙骨碎片中由4个平滑锁定钉组成的结构明显弱于4个锁定螺钉的结构(分别为626 N和981 N)。在月牙骨尺骨尺侧位置上带有平滑锁定钉的结构比在这些位置上带有锁定螺钉的结构要弱(分别为737 N和977 N)。十二指肠软骨下软骨位置的锁定螺钉比这些位置的平滑锁定钉(分别为1,227 N和934 N)和任何其他组合(分别为1,227 N和942 N)强。结论:对于AO C3关节内distal骨远端骨折,尤其是软骨下和月牙骨尺侧骨折,使用光滑的固定钉而不是光滑的固定钉,可以优化结构的稳定性。这可能会对术后康复方案产生影响,并可能限制与使用掌侧锁定板有关的费用。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号