Evaluation of the influence of tilt of optic disc on the measurement of optic disc variables obtained by optical coherence tomography and confocal scanning laser ophthalmoscopy.

机译:评估光盘倾斜度对通过光学相干断层扫描和共焦扫描激光检眼镜获得的光盘变量测量的影响。

获取原文
获取原文并翻译 | 示例
           

摘要

PURPOSE:: To evaluate the relationship between optic disc variables measured by confocal scanning laser ophthalmoscopy (CSLO) and optical coherence tomography (OCT). METHODS:: Fifty-one eyes in 27 non-glaucomatous patients were scanned using a CSLO-TopSStrade mark and an OCT-OCT 3000trade mark. The relationship between the following four optic disc variables-disc area (DA), cup area (CA), neuroretinal rim area (NRRA), and cup-disc area ratio (CDAR)-measured by CSLO and OCT were evaluated, and the relationship between these measurements and optic disc tilt was assessed. Horizontal (HD) and vertical diameters (VD) of the optic disc were measured from disc photographs, and by CSLO and OCT, and ratios (HD/VD) were then compared. RESULTS:: There was a good correlation in DA, CA, and CDAR except NRRA between OCT and CSLO measurements. In eyes with a tilt >/=4 degrees, DA and NRRA were measured larger, and CDAR and CA were measured smaller by OCT than by CSLO. However, in the eyes with a tilt of <4 degrees, no significant differences in measurement of disc variables were observed for the two measurement systems. HD/VD measured from disc photographs was well correlated with those determined by CSLO (r = 0.741, P < 0.0001), however, it correlated poorly with those measured by OCT in whole study eyes (r = 0.410, P = 0.008) and in eyes tilted by >/=4 degrees, respectively (P = 0.280). CONCLUSIONS:: Although CSLO and OCT measurements of the optic disc are highly correlated, discrepancy becomes prominent in eyes with a disc tilted >/=4 degrees. Therefore, analysis of the optic nerve head by OCT must be cautiously interpreted in eyes with a highly tilted disc.
机译:目的::评估通过共聚焦扫描激光检眼镜(CSLO)和光学相干断层扫描(OCT)测量的视盘变量之间的关系。方法:使用CSLO-TopSS商标和OCT-OCT 3000商标对27例非青光眼患者的51只眼睛进行了扫描。评估通过CSLO和OCT测量的以下四个视盘变量-盘面积(DA),杯面积(CA),神经视网膜边缘面积(NRRA)和杯面积比(CDAR)之间的关系,并且该关系在这些测量和视盘倾斜之间评估。从光盘照片中测量视盘的水平(HD)和垂直直径(VD),并通过CSLO和OCT测量,然后比较比率(HD / VD)。结果:除了OCT和CSLO测量之间的NRRA,DA,CA和CDAR的相关性都很好。在OCT大于CSLO的情况下,倾斜> / = 4度的眼睛中DA和NRRA较大,而CDAR和CA较小。但是,在倾斜度小于4度的眼睛中,对于两种测量系统,在椎间盘变量的测量中均未观察到显着差异。从圆盘照片测得的HD / VD与CSLO确定的相关性很好(r = 0.741,P <0.0001),但在整个研究眼睛中,其与OCT测量的相关性很差(r = 0.410,P = 0.008)和眼睛分别倾斜> / = 4度(P = 0.280)。结论:尽管视盘的CSLO和OCT测量值高度相关,但是在视盘倾斜> / = 4度的情况下,眼球之间的差异变得尤为突出。因此,对于高度倾斜的椎间盘,必须谨慎地解释通过OCT对视神经乳头的分析。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号