首页> 外文期刊>Journal of glaucoma >Comparison of a remote operating slit-lamp microscope system with a conventional slit-lamp microscope system for examination of trabeculectomy eyes.
【24h】

Comparison of a remote operating slit-lamp microscope system with a conventional slit-lamp microscope system for examination of trabeculectomy eyes.

机译:远程操作裂隙灯显微镜系统与常规裂隙灯显微镜系统在小梁切除术眼睛检查中的比较。

获取原文
获取原文并翻译 | 示例
           

摘要

To compare 2-dimensional (2D) photo imaging with slit-lamp photo imaging for examination of trabeculectomy eyes, and to compare the accuracy and consistency of examination of trabeculectomy eyes with a remote operating slit-lamp microscope system (referred to as "remote slit lamp" hereafter) and a conventional slit-lamp microscope system (referred to as "slit lamp" hereafter).Thirty-five eyes of 35 patients having a history of trabeculectomy were enrolled in the study that compared 2D photo imaging with slit-lamp photo imaging for the evaluation of trabeculectomy eyes. Five ophthalmologists evaluated the 2D images and the slit-lamp images independently with masking of patient information. Evaluation scores were compared with those provided by a glaucoma specialist using the slit lamp as standard. Fifteen eyes from 15 patients having a history of trabeculectomy were enrolled in the study that investigated the accuracy and consistency of examination of trabeculectomy eyes with a remote slit lamp and a slit lamp.Central anterior chamber depth, bleb height, and bleb extent evaluated by the slit-lamp photo imaging showed significantly higher consistency with the standard than those evaluated by 2D photo imaging (P<0.05). The remote slit lamp showed good consistency of all the evaluated parameters with the slit lamp and the κ scores of all the evaluated parameters were higher than 0.8. The completion time for evaluation with the remote slit lamp and the slit lamp were 247.3±153.5 and 123.5±53.7 seconds, respectively (P<0.001).Slit-lamp photo imaging is a superior method for examination of trabeculectomy eyes compared with 2D photo imaging. The remote slit-lamp system shows similar potential to the slit-lamp system for the evaluation of trabeculectomy eyes, although the evaluation time is much longer.
机译:比较二维(2D)摄影图像和裂隙灯照相成像检查小梁切除术眼睛的情况,并比较使用遥控裂隙灯显微镜系统(称为“远程裂隙术”)对小梁切除术眼睛进行检查的准确性和一致性灯”)和传统的裂隙灯显微镜系统(以下称为“裂隙灯”)。该研究纳入了35名有小梁切除术史的患者的35只眼,该研究将2D照片成像与裂隙灯照片进行了比较影像学评估小梁切除术的眼睛。五名眼科医生在掩盖患者信息的情况下独立评估了2D图像和裂隙灯图像。使用裂隙灯作为标准,将评估分数与青光眼专家提供的分数进行比较。该研究纳入了15例有小梁切除术史的患者中的15只眼,研究了使用远视裂隙灯和裂隙灯检查小梁切除术眼的准确性和一致性,并通过该方法评估了中央前房深度,气泡高度和气泡程度裂隙灯照相成像显示出与标准品相比,与2D照相成像所评估的相比具有更高的一致性(P <0.05)。远程裂隙灯与裂隙灯显示了所有评估参数的良好一致性,并且所有评估参数的κ得分均高于0.8。远程裂隙灯和裂隙灯的评估完成时间分别为247.3±153.5和123.5±53.7秒(P <0.001)。裂隙灯照相成像是检查小梁切除术眼睛的一种优于2D照相成像的方法。尽管评估时间要长得多,但远程裂隙灯系统显示出与裂隙灯系统类似的潜力,可用于评估小梁切除术的眼睛。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号