首页> 外文期刊>Journal of feline medicine and surgery >Measuring statistical agreement between four point of care (POC) lactate meters and a laboratory blood analyzer in cats.
【24h】

Measuring statistical agreement between four point of care (POC) lactate meters and a laboratory blood analyzer in cats.

机译:在猫中测量四个即时护理(POC)乳酸计和实验室血液分析仪之间的统计一致性。

获取原文
获取原文并翻译 | 示例
           

摘要

The use of blood lactate concentrations as a prognostic indicator and therapeutic gauge in feline medicine has been hindered by the inability to obtain values in a timely manner with minimal quantities of blood. Recently, hand-held point-of-care (POC) lactate meters have become commercially available. The objective of this prospective study was to determine if lactate values produced by three commercially available and one medical grade POC meter were in agreement with a laboratory blood analyzer. Blood samples from 47 cats were collected on presentation to an emergency service and processed on four POC meters and a Stat Profile Critical Care Xpress blood analyzer. The results were analyzed using the Bland-Altman method. The blood lactate values produced by the hospital grade POC meter and one of the commercially POC meters were in good agreement with the Critical Care Xpress blood analyzer. Other commercially available POC meters produced acceptable agreement.
机译:由于不能及时地以最少的血液量获取数值,因此阻碍了血液乳酸浓度在猫科动物医学中作为预后指标和治疗指标的障碍。最近,手持式即时护理(POC)乳酸计已在市场上出售。这项前瞻性研究的目的是确定由三台市售和一台医用POC仪产生的乳酸值是否与实验室血液分析仪一致。收集了47只猫的血样,并提供给紧急服务人员,并在四个POC仪表和Stat Profile Critical Care Xpress血液分析仪上进行处理。使用Bland-Altman方法分析结果。医院级POC仪和其中一种商用POC仪产生的血乳酸值与Critical Care Xpress血液分析仪非常吻合。其他可商购的POC表产生了可接受的协议。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号