首页> 外文期刊>Journal of endovascular therapy: an official journal of the International Society of Endovascular Specialists >Suprarenal fixation: effect on blood flow of an endoluminal stent wire across an arterial orifice.
【24h】

Suprarenal fixation: effect on blood flow of an endoluminal stent wire across an arterial orifice.

机译:肾上固定:对腔内支架线穿过动脉孔的血流的影响。

获取原文
获取原文并翻译 | 示例
       

摘要

PURPOSE: To investigate what effect, if any, the presence of a stent wire in front of a renal artery has on the volume flow rate of blood through the renal artery. METHODS: Experimental, numerical, and analytical modeling methods were used to test 4 separate stent wire configurations: a stent wire across the center of an artery orifice, an off-center wire placed at one-quarter the arterial diameter, a V-shaped wire with its vertex at the center, and 2 stent wires at one-third-diameter spacing. RESULTS: For all the configurations studied, the presence of stent wires has a minimal effect on the blood flow rate into an artery of >/=3-mm diameter, with most flow rates decreasing by around 1%. This is true provided that there is no buildup of material on the wire. When material buildup was "encouraged" to occur, then decreases in flow rate of up to 40% were observed. The numerical and analytical methods indicated that the flow rates would, in most cases, decrease by around 3% to 10%. CONCLUSIONS: A bare stent wire in front of a >3-mm-diameter artery decreases the flow rate minimally, providing there is no material on the wire. Although the numerical and analytical methods indicated a greater effect on flow, the approximations required for these 2 methods to obtain meaningful solutions suggest that the experimental results are the most accurate. Nonetheless, the analytical equations provided a useful approximation for determining the effect on blood flow due to the presence of a stent wire.
机译:目的:研究肾脏动脉前部支架线的存在(如果有)对通过肾脏动脉的血流量的影响。方法:使用实验,数值和分析建模方法来测试4种单独的支架线配置:横跨动脉孔中心的支架线,放置在四分之一动脉直径处的偏心线,V形线以其顶点为中心,两根支架线之间的直径为三分之一。结果:对于所研究的所有配置,支架线的存在对流入直径> / = 3-mm的动脉的血流速度的影响最小,大多数血流速度降低了约1%。只要导线上没有材料堆积,这是正确的。当“鼓励”发生材料堆积时,观察到流速降低了40%。数值和分析方法表明,在大多数情况下,流速将降低约3%至10%。结论:在直径大于3毫米的动脉前面的裸露支架线会最小程度地降低流速,前提是线上没有任何物质。尽管数值和分析方法表明对流量的影响更大,但这两种方法获得有意义的解所需的近似值表明实验结果是最准确的。尽管如此,该分析方程式提供了有用的近似值,用于确定由于存在支架线而对血流的影响。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号