Requirements for quantitative analysis of intimal reaction in arteries treated with intraluminal stents.

机译:腔内支架治疗的动脉内膜反应定量分析的要求。

获取原文
获取原文并翻译 | 示例
           

摘要

PURPOSE: To assess the reliability of quantitative histomorphometry measurements for the intimal reaction that occurs after intraluminal stenting in an animal model. METHODS: Two self-expanding intravascular stents, a biodegradable poly-D/L-lactic acid (PLA) stent and a stainless steel Wallstent, were implanted in the common iliac arteries of a beagle dog. After 45 days, the histomorphometry of the stented iliac artery segments was quantitatively measured. The relative standard error (RSE) of the estimate was calculated, and the reliability of measurements for maximal (LDmax) and minimal (LDmin) luminal diameters and internal (IELT) and external (EELT) elastic lamina thicknesses was assessed. RESULTS: The PLA stent required more measurements of variables from a single slide to ensure reliable (RSE<10%) results (1 for LDmax, 1 for LDmin, 8 for IELT, and 4 for EELT) compared to the Wallstent (1 for LDmax, 1 for LDmin, 2 for IELT, and 2 for EELT 2). The measured results were reliable for both stent materials when variables were measured from 2 slides of each segment (proximal/central/distal) of the stent. CONCLUSIONS: We conclude that these preliminary measurements to estimate the reliability of quantitative histomorphometry measurements should be made and reported before final results are given.
机译:目的:评估在动物模型中腔内支架置入后发生的内膜反应的定量组织形态计量学测量的可靠性。方法:在比格犬的总动脉中植入了两个自膨胀血管内支架,可生物降解的聚D / L-乳酸(PLA)支架和不锈钢Wallstent。 45天后,定量测量支架动脉节段的组织形态。计算估计值的相对标准误差(RSE),并评估最大(LDmax)和最小(LDmin)内腔直径以及内部(IELT)和外部(EELT)弹性薄片厚度的测量可靠性。结果:与Wallstent(1个LDmax)相比,PLA支架需要对单个载玻片进行更多的变量测量,以确保可靠的结果(RSE <10%)(LDmax 1个,LDmin 1个,IELT 8个和EELT 4个)。 ,LDmin为1,IELT为2,EELT 2为2)。当从支架的每个部分(近端/中心/远端)的两个载玻片上测量变量时,两种支架材料的测量结果都是可靠的。结论:我们得出结论,在给出最终结果之前,应进行这些初步测量以估计定量组织形态计量学测量的可靠性。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号